13

 

 

 

! Warning 

 

Please don’t connect the anode and cathode with lead of battery directly, it will cause danger. 

 

Please don’t put the battery on fire. It may cause explosion. 

 

Please don’t disassemble the battery privately. 

 

The battery should be taken gently, please don’t strike it with other article. 

9 Repairing and Maintenance 

Attention

 

Before  cleaning  the  device,  make  sure  that  the  equipment  is  switched  off  and  disconnected 

from the power line. 

 

If there is any sign that the ECG cable may be damaged or deteriorated, replace it with a new 

one instead of continuing its application on the patient. 

9.1 Maintenance   

 

Do not open the shell of the instrument avoiding the possible electric shock. Any reparation and 

update of  the instrument must  be operated  by professional being trained and authorized. Only 

repair with the accessories produced by our company. 

 

The instrument guarantees for one year, and do not arbitrary disassembly. 

 

Remove  the  electric  plug  when  there  is  power  cut,  the  instrument  must  be  kept  cool  and  dry 

when it is unused for long time; electrify the instrument every three months. 

 

Inspect and maintain of Lead, electrode. 

1

 

The  state  of  leads  can  be  checked  by  multimeter,  please  check  the  leads  according  to 

following list. If the resistance between electrode and relevant insert needle is less than 100ohm, the 

lead  is  up  to  grade.  Please  check  the  leads  periodically,  because  damage  of  any  lead  will  cause 

wrong waveform, the lead can be cleaned by water and soap, be disinfected with 75% alcohol(please 
don’t dip the lead in the liquid).           

 

Lead symbol 

RF 

C1 

C2 

C3 

C4 

C5 

C6 

Insert needle’s position 

10 

11 

14 

12 

 

2

  Curve and knot of the lead will shorten the leads’ lifespan, so please tidy the leads orderly 

before use, then connect it with ECG machine. 

3)  The  electrode  must  be  reserved  in  proper  way,  after  long  period  use,  its  surface  will  be 

oxidated, which may affect test result, so please change the electrode. 

Содержание 33257

Страница 1: ...Code 33257 CARDIOPOCKET ECG USER MANUAL ...

Страница 2: ... for details Please read the Manual very carefully before using this equipment These instructions describe the operating procedures to be followed strictly Failure to follow these instructions can cause measuring abnormality equipment damage and personal injury The manufacturer is NOT responsible for the safety reliability and performance issues and any monitoring abnormality personal injury and e...

Страница 3: ...rview 2 3 Principle and Caution 3 4 Technical specifications 4 5 Accessories 4 6 Installation 5 7 Operating Guide 8 8 The notice of the battery operating 12 9 Repairing and Maintenance 13 10 Troubleshooting 14 11 Key of Symbols 17 ...

Страница 4: ...ded by our company if the defibrillation time is over 5 s you should use one off chest electrode which can avoid the metal electrode burn patient s skin Please don t use it with electric stimulator at the same time if it is compulsory a professional technician must guide on the scene 1 2 Warnings Explosive hazard DO NOT use the Electrocardiograph in environment with inflammable gas such as some ig...

Страница 5: ...0 05 150Hz frequency response which is the world level also the equipment has advanced digital baseline drifting filter Alternating Current filter and Myo electricity disturbance filter The sampling rate of lead signal is 800Hz the closure frequency of low pass filter machine with guards against mixes folds is 3dB 220Hz The sampling rate is higher than the standard of not lower than 500Hz which ma...

Страница 6: ...tem coordinates the operation of print controlling equipment liquid crystal screen and keyboard driver it is responsible for the profile print driver controlling information demonstration keyboard management and some other control of CMS80 3 2 Caution 3 2 1 Observe the status of the patients and the instrument momentarily 3 2 2 Use the patient cable only to connect the patient and instrument 3 2 3...

Страница 7: ... provided by the factory It may damage this equipment if using the accessories with different models Power adapter Input AC 100 240V 50 60Hz Output DC 9V Attention User can choose the power cord according the request of local power net and also the safe standard Power supply adapter should be accorded with relevant regulation according to IEC60601 1 One lithium rechargeable battery DC 7 4V lithium...

Страница 8: ... 5 6 Installation 6 1 Sketch on each side and accessory Front view ...

Страница 9: ...d V4 V4 Fifth inter costal space at left mid clavicular line V5 Left anterior axillary line at the horizontal lever of V4 V6 Left mid axillary line at the horizontal lever of V4 Clean the positions around to which chest electrodes are to be attached with alcohol lay the ECG medical gel on dia 25cm area around the position and the edge of chest electrode cap press the suction bulb firmly attach the...

Страница 10: ...d skin Attention Tighten the knob after inserting the mainframe patient cable pin in the patient cable socket 6 2 3 Electrode connection definitions and color code Electrode Location Electrode Code Line Color Electrode Color Socket Number Right arm R Black Red 9 Left arm L Black Yellow 10 Left leg F Black Green 11 Right leg RF Black Black 14 Chest 1 V1 C1 White Red 12 Chest 2 V2 C2 White Yellow 1 ...

Страница 11: ...ing paper according to the illustration then clip them in the relevant position in paper fixer 3 Close the paper cabinet cover and protrudes 2cm in the top of the recording paper 7 Operating Guide 7 1 The instrument starts self checking after its turning on In the self checking process 7 2 The instrument will show the following state after self checking under DEMO mode ...

Страница 12: ...ning for example then press the button to switch the relevant information this column is read only The state and other indicator information are all read only Mode Conversion shows the mode conversion of the systemic lead it divides into Auto lead conversion AUTO and Manual lead conversion MANU Leading condition indication it will show NORM when the leads collecting signals normally and OVER when ...

Страница 13: ...tton as followings Press button to return after installing paper 5 Calibrating Press button to print a lmv standard voltage mark in order to observe the current sensitivity directly The LCD will display the following picture when in calibrating process In calibrating process Please Waiting English Interface Ending the Calibrating English Interface ATTENTION The calibrating process is operated auto...

Страница 14: ...trument records the twelve patient cables automatically three seconds ECG signal each patient cable Filter OFF 50Hz 60Hz 50Hz 60Hz totally five types of filter among them 50Hz and 60Hz denote the opening of 35Hz EMG filter ATTENTION Switching on the EMG interference filter will fall the range of the recorded R wave Speed 25mm s 50mm s totally two types of paper skipping speed mode System Manual En...

Страница 15: ...When the battery is full the device can be operated for one hour when the battery be used as power supply An icon of battery will be displayed in the LCD screen of front panel this icon indicates power of battery When the battery is power off the device will turn off automatically this setting is for avoiding permanent damage on battery caused by excessive discharge 3 please charge the battery aft...

Страница 16: ...t guarantees for one year and do not arbitrary disassembly Remove the electric plug when there is power cut the instrument must be kept cool and dry when it is unused for long time electrify the instrument every three months Inspect and maintain of Lead electrode 1 The state of leads can be checked by multimeter please check the leads according to following list If the resistance between electrode...

Страница 17: ...eck battery capacity The instrument has the function of protecting the electro circuit when the battery over discharged 2 Check the AC power voltage The instrument has the function of protecting the electro circuit when its overvoltage 3 Check AC interference check the knob on the plug of the patient cable The instrument has the function of protecting the electro circuit when over loading 10 2 AC ...

Страница 18: ... electrodes and patients body 4 Keep the patient from motion or hyperventilation 5 Check the connection between patient cable and electrodes Please use filter if still having above mentioned interference 10 5 Troubleshooting Trouble Possible Reason Solution Big interference Random waveform 1 It is firm whether the grounding connections 2 The joint of two sides of conductance line is not stable 3 A...

Страница 19: ...nd leader s plug part is not connection enough 5 the connection between patient cable and electrodes is not well 1 The alcohol of use is asked for well 2 Use ethyl alcohol to wash electrode wafer and the skin under the electrode 3 Be charged to the battery 4 The electrode reed tightly clip Baseline drift 1 The batteries are drained 2 Patient moves 1 Be charged to the battery 2 Keep patients quietl...

Страница 20: ...n key power supply acknowledgement shortcut menu Function Key Pause On Function Key System Menu Function Key Print Function Key Calibrate AC power is on no battery The battery solely powering and the battery is full of electricity The battery solely powering and the battery is not full of electricity The battery solely powering and the electric power is going to be used up On charging Danger volta...

Страница 21: ...rts due to production reasons Labor costs and personnel traveling expenses and packaging not included All components subject to wear are not included in the warranty The repair or replacement performed during the warranty period shall not extend the warranty The warranty is void in the following cases repairs performed by unauthorized personnel or with non original spare parts defects caused by ne...

Отзывы: