Requirements for transport and storage:
• No corrosion gas and well-ventilated room.
• Keep the device out of the reach of children.
• Do not store the device in conditions such as direct sunlight, high temperature, humidity, dust or water infiltrations, etc.
• Keep the device away from sloping surfaces, vibrations or shocks.
• Transportation adopts general transportation means or follows the contract requirements. Avoid violent shock, vibration, rain
and snow splash during the process of transportation.
5.4 Pollution-free disposal and recycling
The service life of product is 3 years. If the device exceeds the period of use, it must be discarded. Please contact the manufac-
turer or distributor for more information.
1) The atomizer out of use can be sent back to the manufacturer or distributor for proper recycling.
2) Used parts can be returned to the manufacturer or distributor for disposal, or treated in accordance with relevant laws and
regulations.
Chapter 6
TROUBLESHOOTING
17
ENGLISH
Problems
Reason analysis
Solutions
The device can't startup.
Battery is not well installed.
Check the installation of battery, and reinstall
the batteries.
No atomizing or little atomizing
appears when device working.
Medication cup is not well installed.
Check the installation of medication cup, and
reinstall it.
No medication in medication cup.
Trickle medication into medication cup, re-
member not to exceed its maximum capacity.
Improper medication.
Consult a doctor about whether the medica-
tion is suitable for the device.
The nebulizing sheet is dirty.
Clean medication cup.
There is water around the nozzle
of nebulizer.
Due to temperature differences, the temper-
ature of medication cup surface is relatively
low, medication mist comes in contact with the
nozzle, then condenses into water droplets.
Remove medication cup, pour the water out.
After startup, power indicator
lights about. 1s, then immediate-
ly goes out.
Medication cup is not well installed.
Install medication cup once again.
Medication cup is not loaded with any med-
ication.
Put the medication into medication cup after
consulting your doctor.
After turning on the device, the
power indicator lights once, then
it is out immediately or the de-
vice cannot work normally.
The battery had run down.
Replace the batteries immediately.
Nebulizer, doesn't automatically
shutdown when medication is
used up.
Medication may generate bubbles in medica-
tion cup.
Press “ON/OFF” button to turn off the device,
and clear up the bubbles.
Medication may be stuck on the nebulizing
sheet.
Press “ON/OFF” button to turn off the device,
and clean medication cup.
The electrodes contacting with the medication
cup may be dirty.
Press “ON/OFF” button to turn off the device,
and clean the electrodes.
If the device still can't work normally alter using all methods above, please contact our after-sales service.