background image

20

ENGLISH

Recommended separation distances between portable and mobile RF communications equipment and the Medical 

28066 / NE-M01

The “28066 / NE-M01” is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. 

The customer or the user of the Medical 28066 / NE-M01 can help prevent electromagnetic interference by maintaining a 

minimum distance between portable and mobile RF communications equipment (transmitters) and the “28066 / NE-M01” as 

recommended below, according to the maximum output power of the communications equipment.

Rated maximum output 

power of transmitter (W)

Separation distance according to frequency of transmitter (m)

150KHz to 80MHz

80MHz to 800MHz

800MHz to 2,5GHz 

0,01

0.12

0.12

0.23

0,1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation distanced in meters (m) can 

be estimated using the equation applicable to the frequency of the transmitter, where 

P

 is the maximum output power rating of 

the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection 

from structures, objects and people.

3.5

E

 

1

√P

d=

 7

E

 

1

√P

d=

3.5

V

 

1

√P

d=

Recommended separation distances between portable and mobile RF communications equipment and the Medical 

28066 / NE-M01

Disposal: The product must not be disposed of along with other domestic waste. The users must dispose of this equip-

ment by bringing it  to a specific recycling point for electric and electronic equipment.

For further information on recycling points contact the local authorities, the local recycling center or the shop where the 

product was purchased. If the equipment is not disposed of correctly, fines or penalties may be applied in accordance 

with the national legislation and regulations. 

GIMA WARRANTY CONDITIONS

Congratulations for purchasing a GIMA product. This product meets high qualitative standards both as regards the material and 

the production. The warranty is valid for 12 months from the date of supply of GIMA.

During the period of validity of the warranty, GIMA will repair and/or replace free of charge all the defected parts due to production 

reasons. Labor costs and personnel traveling expenses and packaging not included. All components subject to wear are not 

included in the warranty. The repair or replacement performed during the warranty period shall not extend the warranty.

The warranty is void in the following cases: repairs performed by unauthorized personnel or with non-original spare parts, defects 

caused by negligence or incorrect use. GIMA cannot be held responsible for malfunctioning on electronic devices or software due 

to outside agents such as: voltage changes, electro-magnetic fields, radio interferences, etc. 

The warranty is void if the above regulations are not observed and if the serial code (if available) has been removed, cancelled 

or changed. The defected products must be returned only to the dealer the product was purchased from. Products sent to GIMA 

will be rejected.

a: Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, 

amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the 

electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the meas-

ured field strength in the location in which the 28066 / NE-M01 is used exceeds the applicable RF compliance level above, the 

28066 / NE-M01 should be observed to verify normal operation. If abnormal performance is observed, additional measures may 

be necessary, such as reorienting or relocating the 28066 / NE-M01.

Содержание 28066

Страница 1: ...TTENTION The operators must carefully read and completely understand the present manual before using the product AVIS Les op rateurs doivent lire et bien comprendre ce manuel avant d utiliser le produ...

Страница 2: ...thetic mixture Otherwise it may cause personal injury Do not wrap the power cord around children s neck or other sensitive areas or it may cause asphyxia The mask of this equipment is made of PVC mate...

Страница 3: ...batteries is DC3V 2 3 Operational environment Temperature 5 C 40 C Humidity 15 90 Atmospheric pressure 700hPa 1060hPa Attention This product is not suitable for use in strong electromagnetic interfer...

Страница 4: ...cup to the main unit be sure to install it properly until clasp sound is heard Otherwise it may cause electrode conduction failure then the device cannot atomize normally 2 Remove medication cup from...

Страница 5: ...ery life will be affected Remove the batteries if the device won t be used for long time 4 2 Operations for treatment use Preparations before use 1 Remove medication cup please clean and disin fect it...

Страница 6: ...the cup lid and discard residual medication 3 Add 75 ethanol solution in medication cup cover the cup lid then leave for at least 10mins it is advisable to gently shake it for better disinfection 4 Im...

Страница 7: ...medication cup Trickle medication into medication cup re member not to exceed its maximum capacity Improper medication Consult a doctor about whether the medica tion is suitable for the device The ne...

Страница 8: ...ode Keep away from sunlight 55 C 40 C Store between 40 and 55 C Lot number Fragile handle with care Date of manufacture Serial number Follow instructions for use Medical Device complies with Directive...

Страница 9: ...4 2 Air Discharge 2 kV 4 kV 8 kV 15 kV contact 8 kV air Air Discharge 2 kV 4kV 8 kV 15 kV contact 8 kV air Floors should be wood concrete or ce ramic tile If floor is covered with synthetic material...

Страница 10: ...ct was purchased If the equipment is not disposed of correctly fines or penalties may be applied in accordance with the national legislation and regulations GIMA WARRANTY CONDITIONS Congratulations fo...