20
ENGLISH
Recommended separation distances between portable and mobile RF communications equipment and the Medical
28066 / NE-M01
The “28066 / NE-M01” is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
The customer or the user of the Medical 28066 / NE-M01 can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment (transmitters) and the “28066 / NE-M01” as
recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter (W)
Separation distance according to frequency of transmitter (m)
150KHz to 80MHz
80MHz to 800MHz
800MHz to 2,5GHz
0,01
0.12
0.12
0.23
0,1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distanced in meters (m) can
be estimated using the equation applicable to the frequency of the transmitter, where
P
is the maximum output power rating of
the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
3.5
E
1
√P
d=
7
E
1
√P
d=
3.5
V
1
√P
d=
Recommended separation distances between portable and mobile RF communications equipment and the Medical
28066 / NE-M01
Disposal: The product must not be disposed of along with other domestic waste. The users must dispose of this equip-
ment by bringing it to a specific recycling point for electric and electronic equipment.
For further information on recycling points contact the local authorities, the local recycling center or the shop where the
product was purchased. If the equipment is not disposed of correctly, fines or penalties may be applied in accordance
with the national legislation and regulations.
GIMA WARRANTY CONDITIONS
Congratulations for purchasing a GIMA product. This product meets high qualitative standards both as regards the material and
the production. The warranty is valid for 12 months from the date of supply of GIMA.
During the period of validity of the warranty, GIMA will repair and/or replace free of charge all the defected parts due to production
reasons. Labor costs and personnel traveling expenses and packaging not included. All components subject to wear are not
included in the warranty. The repair or replacement performed during the warranty period shall not extend the warranty.
The warranty is void in the following cases: repairs performed by unauthorized personnel or with non-original spare parts, defects
caused by negligence or incorrect use. GIMA cannot be held responsible for malfunctioning on electronic devices or software due
to outside agents such as: voltage changes, electro-magnetic fields, radio interferences, etc.
The warranty is void if the above regulations are not observed and if the serial code (if available) has been removed, cancelled
or changed. The defected products must be returned only to the dealer the product was purchased from. Products sent to GIMA
will be rejected.
a: Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the meas-
ured field strength in the location in which the 28066 / NE-M01 is used exceeds the applicable RF compliance level above, the
28066 / NE-M01 should be observed to verify normal operation. If abnormal performance is observed, additional measures may
be necessary, such as reorienting or relocating the 28066 / NE-M01.