Volume Setting

Do not adjust the volume without squeezing both parts of the pipetting handle

 

.

The volume of liquid to be aspirated is set 
using the volume adjustment wheel

 

.

With the pipetting handle fully squeezed, 

the volume is set by slowly turning the 

volume adjustment wheel  to reach the 

required setting. 
Never try to force the volume outside of 

the volume range permitted by the device.

Tip Fitting

For optimum performance, use of PIPETMAN® DIAMOND Tips with PLATEMASTER is 

strongly recommended. Plastic tips are for a single use and should not be cleaned for 

reuse.

To fit PIPETMAN® DIAMOND Tips on PLATEMASTER:

• 

Raise the pipetting head vertically 
using the pipetting handle. Move the 
pipetting head 

 

to the left side.

• 

Slide the right drawer out 

 

and place 

a new tip rack on the back position. 
The four locations have been specially 
designed for PIPETMAN DIAMOND 
Tips TIPACK™. 

Slots embedded in 

the drawers ensure that tip racks 

securely fit into all locations. 

• 

Use the pipetting handle 

 

to position 

the pipetting head 

 to the right and 

lower it far enough so that a contact is 
made with the tips in the rack. 

• 

Press down on the tip fitting handle

 

 

with moderate pressure. The tip ejector must be fully expanded to ensure that the 
tips are fully attached.

 

.

• 

Move the pipetting head

 

 up and the right drawer

 

 

back.

The device is now ready to pipette liquid. 

5

3

1

4

2

Pipetting

Place a vessel containing the solution to be pipetted in the front location of the right 

drawer. The solution should be under the pipetting head.

Liquids containing protein solutions and organic solvents can leave a film of liquid 

on the inside wall of the tip; pre-rinse the tip to minimize any errors related to this 

phenomenon. Pre-rinsing consists of aspirating the first volume of liquid and then 

dispensing it back into the same vessel (or to waste). Subsequent volumes pipetted 

will have levels of accuracy and precision within specifications. As recommended by 

the ISO 8655 standard, pre-rinse to stabilize dead volume inside the pipetting head.

ASPIRATION

• 

Gently squeeze the pipetting handle

 

 

to the first stop

 (this corresponds 

to the set volume of liquid).

• 

Use the pipetting handle 

 

to lower 

the pipetting head 

 

to immerse the 

tips in the liquid (see immersion depth 
table below). 

It is important that the 

tips do not touch the bottoms of the 

wells.

• 

Slowly retract the pipetting handle to 

aspirate the liquid into the tips.

• 

Move the pipetting head fully up.

To simplify this operation, the height 
adjuster

 

 

can be set by turning the upper 

screw so that the pipetting head may only be 
moved down to a position slightly above the 
bottom.

MODEL

IMMERSION DEPTH 

(MILLIMETERS)

WAIT TIME 

(SECONDS)

P20

2-3

1

P220

2-4

1

Check the tip-ejector when fitting tips. If the tips have shifted this 

implies that they have not been fitted correctly.

The maximum 

immersion depth 

of the tips can be 

adjusted using a 

height adjuster.

O

PE
R

A

TIO

N

O

PE

R

A

TIO

N

9

8

PLATEMASTER®

 

|

 

USER’S GUIDE

OPERATION

 

PLATEMASTER®

Содержание PLATEMASTER P20

Страница 1: ...User s Guide EN...

Страница 2: ...g master mixes for PCR qPCR Simultaneous processing of nearly all reaction steps in microplates coating starting stopping washing etc for applications such as ELISA Addition and removal of culture med...

Страница 3: ...ling of the pipetting head 1 transport box and packing material Gilson original certificate of conformity PLATEMASTER P20 part number F110761 contains PLATEMASTER P20 User s guide 3 height adjusters 1...

Страница 4: ...increments 2 to 220 L Volume can be adjusted in 1 L increments Temperature Range Recommended working temperature 20 C to 25 C 68 F to 77 F Dimensions 300 x 450 x 400 mm 12 x 18 x 16 Inch W x L x H Wei...

Страница 5: ...a vessel containing the solution to be pipetted in the front location of the right drawer The solution should be under the pipetting head Liquids containing protein solutions and organic solvents can...

Страница 6: ...icroplate and dispense the first 96 wells STEP 2 Move the microplate on to the second position of the 384 well plate adapter using the blue wheel Aspirate the solution Move the pipetting head over the...

Страница 7: ...id sample or reagent Each new tip should be pre rinsed with the liquid to be pipetted Do not pipette liquids having temperatures above 70 C or below 4 C PLATEMASTER can be used between 4 C and 40 C bu...

Страница 8: ...nium Use of the Lubrication Box In case of intensive use between each maintenance interval and to avoid bad tip fitting it is recommended to lightly lubricate the external O ring using the lubrication...

Страница 9: ...es in Microliters Chapter 8 WARRANTY Gilson warrants this device against defects in material under normal use and service for a period of 12 months from the date of purchase This warranty shall not ap...

Страница 10: ...NOTES NOTES Notes Notes 19 18 PLATEMASTER USER S GUIDE Notes PLATEMASTER...

Страница 11: ...ed except as allowed under copyright laws www gilson com contactus Gilson Inc 3000 Parmenter Street PO Box 620027 Middleton WI 53562 USA Tel 608 836 1551 or 800 445 7661 F 608 831 4451 Gilson S A S 19...

Отзывы: