8
PIPETMAN
®
P
in the liquid (see immersion depth table).
Release the push-button slowly and
smoothly (to
top
position) to aspirate the set
volume of liquid.
Wait one second (time depends on model,
see table); then withdraw the pipette-tip
from the liquid.
You may wipe any droplets away from the
outside of the tip using a medical wipe,
however if you do so
take care to avoid
touching the tip’s orifice
.
4) Dispense.
Place the end of the tip against the inside
wall of the recipient vessel (at an angle of
10° to 40°).
Press the push-button slowly and smoothly
to the
first stop
.
Wait for at least a second; then press the
push-button to the
second stop
to expel
any residual liquid from the tip.
Keep the push-button pressed fully down
and (while removing the pipette) draw the
tip along the inside surface of the vessel.
Release the push-button, smoothly.
Top
1st Stop
2nd Stop