23
ENGLISH
Gilson declares that its manufactured pipettes
comply with the requirements of the ISO 8655
Standard, by type testing. The adjustment is
carried out under strictly defined and monitored
conditions (ISO 8655-6):
• Basis of adjustment, Ex.
• Reference temperature, 20 °C
• Relative humidity, 50%
• Barometric pressure, 101 kPa
• Use of distilled water grade 3 (ISO 3696)
• Ten measurements for each test volume, which
are Nominal Volume, 50% of Nominal Volume,
and the minimum or 10% of Nominal Volume.
Maximum Permissible Errors
Model
Volume
(µL)
Gilson
ISO 8655
Systematic Random Systematic Random
error (µL) error (µL) error (µL) error (µL)
P2
Min
0.2
± 0.024
압
0.012
± 0.08
압
0.04
(F144801)
0.5
± 0.025
압
0.012
± 0.08
압
0.04
Max.
2
± 0.030
압
0.014
± 0.08
압
0.04
P10
Min.
1
± 0.025
압
0.012
± 0.12
압
0.08
(F144802)
5
± 0.075
압
0.030
± 0.12
압
0.08
Max
10
± 0.100
압
0.040
± 0.12
압
0.08
Min.
2
± 0.10
압
0.03
± 0.20
압
0.10
P20
5
± 0.10
압
0.04
± 0.20
압
0.10
(F123600)
10
± 0.10
압
0.05
± 0.20
압
0.10
Max.
20
± 0.20
압
0.06
± 0.20
압
0.10
P100
Min.
20
± 0.35
압
0.10
± 0.80
압
0.30
(F123615)
50
± 0.40
압
0.12
± 0.80
압
0.30
Max.
100
± 0.80
압
0.15
± 0.80
압
0.30
P200
Min.
50
± 0.50
압
0.20
± 1.60
압
0.60
(F123601)
100
± 0.80
압
0.25
± 1.60
압
0.60
Max.
200
± 1.60
압
0.30
± 1.60
압
0.60
P1000
Min.
200
± 3
압
0.6
± 8
압
3.0
(F123602)
500
± 4
압
1.0
± 8
압
3.0
Max. 1000
± 8
압
1.5
± 8
압
3.0
P5000
Min. 1000
± 12
압
3
± 40
압
15
(F123603)
2000
± 12
압
5
± 40
압
15
Max. 5000
± 30
압
8
± 40
압
15
Min. 1 mL
± 30
압
6
± 60
압
30
P10ml
2 mL
± 30
압
6
± 60
압
30
(F161201)
5 mL
± 40
압
10
± 60
압
30
Max. 10 mL
± 60
압
16
± 60
압
30
(Reference)
(Accuracy) (Precision)(Accuracy)(Precision)
Содержание PIPETMAN P
Страница 1: ...PIPETMAN P ENGLISH ...
Страница 24: ...24 PIPETMAN P 15 SPARE PARTS X E E A C H S1 S2 P2 and P10 only ...
Страница 26: ...26 PIPETMAN P NOTES ...
Страница 27: ...27 ENGLISH ...
Страница 29: ...PIPETMAN P FRANÇAIS ...
Страница 55: ...27 FRANÇAIS NOTES ...
Страница 57: ...PIPETMAN P DEUTSCH ...