PIPETMAN
®
P
Tip-ejector and Tip-holder
These parts must be changed, if they are
accidentally damaged or attacked chemically.
You should also remove these parts for cleaning
or decontamination purposes. A cloth
dampened with ethanol may be used to
routinely clean the outside of the pipette.
Changing the Tip-ejector
1) To remove the tip-ejector, keep the tip-
ejector button depressed and pull down on
the flanged upper part of the tip-ejector with
the other hand (moderate force is required).
2) To refit the tip-ejector, keep the tip-ejector
button depressed, slide the end of the tip-
ejector over the end of the tip-holder and
push the plastic end of the tip-ejector back
into the body of the pipette until it is gripped
firmly by the tip-ejector rod (metal).
Changing the Tip-holder
1) Remove the tip-ejector (see above).
2) Unscrew the connecting nut by turning it
counterclockwise, by hand.
3) Carefully separate the lower and upper parts.
4) Remove the piston assembly, O-ring and
seal.
5) Clean, autoclave, or replace the tip-holder.
6) Reassemble the pipette (refer to the figure).
7) Tighten the connecting nut (turn by hand,
clockwise).
8) Refit the tip-ejector (see above).
Servicing the Piston Assembly
You may remove the piston assembly for
cleaning purposes only. If the piston assembly
is changed, the pipette must be adjusted and
calibrated.
Содержание PIPETMAN P
Страница 1: ...PIPETMAN P ENGLISH ...
Страница 24: ...24 PIPETMAN P 15 SPARE PARTS X E E A C H S1 S2 P2 and P10 only ...
Страница 26: ...26 PIPETMAN P NOTES ...
Страница 27: ...27 ENGLISH ...
Страница 29: ...PIPETMAN P FRANÇAIS ...
Страница 55: ...27 FRANÇAIS NOTES ...
Страница 57: ...PIPETMAN P DEUTSCH ...