21
Multichannel - Lower Part (Volumetric module)
The following components
only
can be immersed in a
cleaning solution: tip-ejector, ejector locks and ejector
spacer.
Remove the tip-ejector and the ejector spacer.
Immerse the tip-ejector, ejector locks and ejector
spacer in the cleaning solution or wipe them with
a soft-cloth or lint-free tissue impregnated with the
cleaning solution.
Rinse the components with distilled water.
Wipe the entire pipette with a soft cloth or lint-free
tissue impregnated with the cleaning solution.
Wipe it with distilled water.
Leave the parts to dry by evaporation or wipe them
with a clean soft-cloth or lint-free tissue.
Refit the tip-ejector as described in page 16 (Chang-
ing the tip-ejector).
13 - sPecIFIcatIons
PIPETMAN Neo is a high quality pipette that offers ex-
cellent accuracy and precision. The figures given in the
“Gilson Maximum Permissible Errors” table (page 22)
were obtained using PIPETMAN DIAMOND Tips. These
figures are guaranteed only when genuine PIPETMAN
DIAMOND Tips are used.
Each pipette is inspected and validated by qualified
technicians in accordance with the Gilson Quality System.
Gilson declares that its manufactured pipettes comply
with the requirements of the ISO 8655 standard, by
type testing.
The adjustment is carried out under strictly defined and
monitored conditions (ISO 8655-6).