background image

Gilson Company, Inc. 

Micro-Deval Apparatus: MD-2000, MD-2000C & MD-2000F

Page 7

•  Close the doors properly to insure the safety interlock 

is active and press the Green <START> button. The 
jars will turn and the controller will count down to zero 
while the revolutions counter counts up to the setpoint. 
Power to the motor will then cut off.

  7.0  OPERATING INSTRUCTIONS:

The MD-2000 is now ready to perform a test.  

For additional tests with the same time or number of 
revolutions, simply press the green <START> button.

NOTE: The counter will retain the last number 
displayed if power is interrupted. The instrument 
can resume a test after a power outage by pressing 
the green <START> button.

7.1  Performing a Test:

To perform a test with the MD-2000, first refer to the test 
protocol in the governing specifications and prepare the 
aggregate specimen as specified.

1.  Load the prepared sample into the abrasion jar 

along with the specified amounts of abrasive charge 
(stainless steel balls) and water.

2.  Carefully clean and dry off the rim of the jar and the 

lid gasket.

3.  Place the lid on the jar and secure with the threaded 

locking knob.

NOTE: Excessive tightening of the knob may result 
in deformation of components. Tighten only until 
snug and free from leaks.

4.  Invert the jar and check for leaks.

5.  Place the jar on its side on the roller set with the clamping 

end facing the optical encoder (see Figure 4). 

6.  Check that timer or revolutions mode is selected 

properly and that settings are correct. If necessary, 
input new settings as noted in previous section.

7.  Check that there are no foreign objects or obstructions 

in the jar chambers.

8.  Close and secure the cabinet doors; first the bottom 

then the top, latching door.

9.  Press the green <START> button. Depending on the 

mode selected, the timer or the revolutions counter 
will then control the duration of the test and stop the 
machine automatically at completion.

NOTE: Pressing the large, red <STOP> button 
located on the front panel will cut power to the 
motor and rotation will cease. Pressing the green 
<START> button will resume operation.

 WARNING!

It is possible for pressure to build inside the sealed 
jars during testing. Use caution when removing the 
jar lid after completion of test.

  8.0  MAINTENANCE:

 WARNING!

Always disconnect and lockout the power supply to 
this machine before performing any maintenance 
or repair.

No normal maintenance procedures require access to 
the cabinet interior. There are no lubrication points on 
the instrument.

Upper Level 

Optical Encoder Sensor

Figure 4

Lower Level 

Optical Encoder Sensor

Содержание MD-2000

Страница 1: ...eval Apparatus MD 2000 MD 2000C MD 2000F Rev 07 17 2017 PHONE 800 444 1508 P O Box 200 Lewis Center Ohio 43035 0200 FAX 800 255 5314 740 548 7298 E mail customerservice gilsonco com Website www global...

Страница 2: ...eory of Operation 3 3 0 Principle of Operation 3 4 0 Unpacking Set Up 3 5 0 Controller Operation 4 5 1 Mode Selection 4 6 0 Initial Function Check 6 7 0 Operating Instructions 7 7 1 Performing a Test...

Страница 3: ...via an optical encoder until they reach zero at which point the unit will stop 4 0 UNPACKING SET UP NOTE Save all packaging materials in case instrument return is required The MD 2000 instrument is sh...

Страница 4: ...preset number of jar revolutions is complete If power is lost and reapplied this screen will reappear Timer Mode Screen This screen appears when A is selected from the Power On screen and will be disp...

Страница 5: ...from single digits to tens hundreds then thousands Once the desired value is displayed press OK again to enter the new value The next two lines are current revolutions for the top and bottom jars upda...

Страница 6: ...ere are no jars on the rollers After about 30 seconds depress the red STOP button The operation of the MD 2000 may now be checked with the jars in place The MD 2000 jars steel balls and magnetic picku...

Страница 7: ...n deformation of components Tighten only until snug and free from leaks 4 Invert the jar and check for leaks 5 Placethejaronitssideontherollersetwiththeclamping end facing the optical encoder see Figu...

Страница 8: ...g clip on the master link by spreading the legs apart and sliding the clip off The link may now be separated and the chain can be removed 8 3 Roller Replacement Replacement idler roller is WGMD ROLLER...

Страница 9: ...Gilson Company Inc Micro Deval Apparatus MD 2000 MD 2000C MD 2000F Page 9 10 0 PARTS DIAGRAM Figure 5 Top Cover Left Top Dustshield Panel Left Mid Dustshield Panel Bottom Door...

Отзывы: