CENTRY™ 117 MiCRoCENTRifugE 

| uSER’S guiDE

oPERATioN 

|

 CENTRY™ 117 MiCRoCENTRifugE

Manual Lid Release

in the event of a power failure or malfunction, it may be 
necessary to open the lid manually.
1.  Disconnect the power cord from the wall socket.
2.  Remove the plastic plug, located on the left side of the unit.
3.  Pull the wire (attached to the plug) to open the lid lock 

manually.

WARNING

Do not attempt to override the lid lock mechanism. 
Doing so is dangerous and could damage the CENTRY 117 
Microcentrifuge.

Speed Selection

The speed (rpm or g-force) can 
be selected from 1,000 to 13,500 
rpm or from 100 to 16,800 x g by 
touching the speed display and 
adjusting with the up and down 
arrow buttons.Display of rpm/rcf 
can be toggled with the rPM/RCf 
button on the display.

Load the Rotor

Tubes to be loaded should be filled equally 
by eye. The difference in the weight 
between the tubes should not exceed 0.1 
gram. Tubes should always be loaded so 
that there is equal spacing between all 
tubes. one or two additional loaded tubes 
may need to be added to achieve this.

Figure 1

MyTouch™ Control Panel Layout

Figure 2

Balanced Rotor Load

Figure 3

Incorrect Rotor Load 

Содержание Centry 117

Страница 1: ...User s Guide EN Centry 117 Microcentrifuge...

Страница 2: ...Trademarks All product and company names are trademarks or registered trademarks of their respective holders Use of the trademark s in this document does not imply any affiliation with or endorsement...

Страница 3: ...117 Microcentrifuge USER S GUIDE TABLE OF CONTENTS Introduction 4 Specifications 4 General Safety Precautions 5 Installation 6 Operation 7 Maintenance 10 Troubleshooting 13 Ordering Information 15 War...

Страница 4: ...than 1 5 mL The microcentrifuge reaches speeds of up to 13 500 rpm 16 800 x g Specifications CENTRY 117 Microcentrifuge Dimensions W x D x H 22 86 x 30 48 x 19 7 cm 9 x 12 x 7 75 in Maximum Speed 13 5...

Страница 5: ...erly The microcentrifuge is partially dismantled Service has been attempted by unauthorized or unqualified personnel The rotor has not been installed securely on the motor shaft Rotors and accessories...

Страница 6: ...r damage must be filed within 48 hours with the carrier The accessories and instruction manual supplied with the CENTRY 117 Microcentrifuge should be kept with the near its place of installation Insta...

Страница 7: ...age the rotors accessories or bowl of the unit Attach Rotor Lid After the rotor has been properly secured and loaded attach the rotor lid to the rotor Always use the rotor lid for safety and to allow...

Страница 8: ...fuge Speed Selection The speed rpm or g force can be selected from 1 000 to 13 500 rpm or from 100 to 16 800 x g by touching the speed display and adjusting with the up and down arrow buttons Display...

Страница 9: ...ifuge will run until manually stopped To start a run press the Start Stop button When the preselected time expires the CENTRY 117 Microcentrifuge will stop automatically Start Stop a Manual Run The CE...

Страница 10: ...Remove the rotor securing screw from the motor shaft by turning the screw counterclockwise using the rotor wrench 2 Lift the rotor directly upward in a straight vertical motion ROTOR INSTALLATION 1 M...

Страница 11: ...chamber rotor and rotor accessories clean All parts should be wiped down periodically with a soft cloth For more thorough cleaning use a neutral cleaning agent pH between 6 and 8 applied with a soft c...

Страница 12: ...on the rear panel 3 Open the fuse holder drawer by inserting a small screwdriver under the tab and prying it open 4 Remove the innermost operative fuse from its retaining tabs and replace the fuse if...

Страница 13: ...rder to centrifuge liquids with a higher density it is necessary to reduce the speed Failure to reduce the speed may result in damage to the rotor and microcentrifuge The revised maximum speed can be...

Страница 14: ...ot closed correctly Close lid correctly No speed or time has been selected Set speed and or time ERROR LABEL CAUSE 1 Lid Open Trying to start when lid is open 2 Lid Unlock Trying start when lid lock n...

Страница 15: ...th COMBI Rotor 230v AU Cord Warranty Gilson warrants this instrument against defects in material under normal use and service for two years from the date of purchase This warranty is valid only if the...

Страница 16: ...ithout notification errors excepted Reproduction adaptation or translation without prior written permission is prohibited except as allowed under copyright laws www gilson com contactus LT318597 01 20...

Отзывы: