MAiNTENANCE
|
CENTRY™ 117 MiCRoCENTRifugE
Maintenance
Rotor Maintenance
The rotor should be cleaned regularly. Thorough cleaning must
be performed when spinning samples containing phenol or
phenol chloroform. Periodically inspect the rotor for dents,
dings, scratches, discoloration, and cracks. if any damage to
the rotor is found, discontinue use of the rotor immediately
and replace.
ROTOR REMOVAL
The rotor can be removed for cleaning.
1. Remove the rotor securing screw from the motor shaft
by turning the screw counterclockwise, using the rotor
wrench.
2. Lift the rotor directly upward in a straight vertical motion.
ROTOR INSTALLATION
1. Make sure that the motor shaft and rotor mounting hole are
clean.
2. Place the rotor on the motor shaft.
3. Reinstall the rotor securing screw on the motor shaft by
turning it clockwise.
4. Hold the rotor with one hand and tighten the rotor securing
screw, using the rotor wrench.
PCR HOLDER REMOVAL
Locate the tabs on each of the two
holders and pull them inward towards
the center of the rotor using a pair of
pliers.
Figure 4
PCR Holders