Serial RS232 connection (COM1)
To make the serial connection, use a shielded cable, taking care to connect the shield to
one of the two ends. If the cable has more wires than those used, connect the free wires
to the shield
The serial connection cable must have a maximum length of 15 meters (EIA RS-232-C
standards), beyond which it is necessary to use the RS485 interface of the instrument.
The cable must not be channeled with other cables (e.g. outputs connected to contactors
or power cables), but must possibly follow its own path.
J1 5 pin female SERIAL
1 GND + shield
2 RX RS232
3
4
Connector J1 frontal view
5 TX RS232
Loose connector rear view
Connect the shield of the cable on one side only!
Serial RS485 connection (COM2)
The connection cable needs to be of a suitable type for serial RS422/RS485 communication
with 1 twisted pair wires for RS485 and relative shielding.
The cable must not be channeled with other cables (e.g. outputs connected to contactors or
power cables), but must possibly follow its own path.
J1 5 pin female SERIAL
1 Shield
2
3 TX- / RX- RS485
4 TX+ / RX+ RS485
Connector J1 frontal view
5
Loose connector rear view
Connect the shield of the cable on one side only!
Connection summary
Below is a summary of the connections to be made on the connectors.
No.
J1
5 pin female
(serial)
J2
5 pin male
(load cell)
J3
4 pin female
(remote input)
1
GND + Shield
Power
Input autotare
2
RX RS232
3
TX- / RX- RS485
Signal -
4
TX+ / RX+ RS485 Power supply -
5
TX RS232
Shield
The (J3-4) has to be connected through button to the autotare input (J3-1).
!
!