Tare of the weight
During the calibration phase, the display shows the weight intermittently with the word "CAL".
During the linearization phase the display shows the weight intermittently with the word "LIN n" where instead
of n there is the number of the point to be set (from 1 to 5).
Zero calibration
Perform the operation on unloaded scale but complete with tare. The visualized
weight must be zeroed. This operation can be repeated several times.
Full scale calibration
Before carrying out the operation, load the sample weight on the scale; the display
shows the detected value to be calibrated.
If the value set is higher than the resolution offered by the instrument, it is not
accepted. It is always possible to repeat the calibration operations.
If the instrument is switched off without exiting the set-up menu, the programming performed
will not be stored.
Performing a zero or full scale calibration or modifying the capacity or division value, any previ-
ously saved linearization points are lost!
!
Pagina –
page
–
Seite
26