– Display in GKL 100 at 25 °C, setpoint = 1413 µS/cm
– Display = 1388 µS/cm
–
S(L
= setpoint / display value
–
S(L
= 1413 µS/cm / 1388 µS/cm = 1.018
NOTE
The most precise results can be achieved when the control solution temperature is ad-
justed to 25 °C. With different temperatures, the table value of the corresponding tem-
perature must be used as a setpoint.
_
Outcome of an action
The changed value is saved and the
Configuration
menu is closed.
NOTE
If the product is switched off without saving the configuration, the last save value is re-
produced on the next start-up of the product.