![GHM G 1410 Скачать руководство пользователя страница 19](http://html1.mh-extra.com/html/ghm/g-1410/g-1410_operating-manual_2220426019.webp)
G 1410
Operation | 7
B-H87.0.01.DB200-2.0
19 / 24
7.2.4 Adjustment of the measuring input
_
Description
_
The temperature input can be adjusted with the zero point correction and the gradient
correction. If an adjustment is made, you change the pre-adjusted factory settings.
This is signalled with the
T.OF
,
T.SL
or
S(L
display text when the product is switched on.
The standard settings of the zero point value and the gradient value of the temperat-
ure input is
0.00
. The standard setting of the gradient value of the conductivity value is
1.000
. It signals that no correction is made.
In order to adjust the product, you must first open the
Adjustment
menu. The menu is
opened as shown in the illustration.
_
Prerequisites
– Sufficiently full batteries are inserted in the product.
– The product is switched off.
– Ice water, regulated precision water baths or a water bath with a reference meas-
urement are available as a reference.
_
Instruction
1. Press and hold the
Down button
.
2. Press the
On/Off button
to switch on the product and open the
Configuration
menu. Release the
Down button
. The display shows the first parameter.
3. By briefly pressing the
Function key
, you can scroll through the parameters. Select
the parameter you would like to configure.
4. When you have selected the desired parameter, change the parameter to the de-
sired value with the
Up button
and the
Down button
.
5. In order to save the new parameter value, press and hold the
Function key
for
longer than 1 second.
_
Representation
Call up menu
Hold
Release
_
Outcome of an action
The
Configuration
menu is closed after the last parameter.
NOTE
If the product is switched off without saving the configuration, the last save value is re-
produced on the next start-up of the product.
7.2.5 Configuring parameters of the adjustment menu
_
Description
The following representation shows the available parameters and various configura-
tion options.
_
Prerequisites
The
Adjustment
Adjustment of the measuring input [
_
Instruction
1. Select the desired parameter you would like to configure.
2. Adjust the desired configuration in the selected parameter with the
Up button
and
Down button
.
3. The available configuration options are listed for each parameter in the following
representation.