![GHM DeltaOHM HD51.3D Series Скачать руководство пользователя страница 30](http://html1.mh-extra.com/html/ghm/deltaohm-hd51-3d-series/deltaohm-hd51-3d-series_operating-manual_2220418030.webp)
HD51.3D…
- 30 -
V2.0
Command
Reply
Description
CAHn
&|
Associates full scale of wind speed analog
output to:
5 m/s if n=0
50 m/s if n=9
10 m/s if n=1
55 m/s if n=10
15 m/s if n=2
60 m/s if n=11
20 m/s if n=3
65 m/s if n=12
25 m/s if n=4
70 m/s if n=13
30 m/s if n=5
75 m/s if n=14
35 m/s if n=6
80 m/s if n=15
40 m/s if n=7
85 m/s if n=16
45 m/s if n=8
90 m/s if n=17
Default
: 75 m/s (n=14)
RAH
& n|
Reads the value corresponding to the wind
speed analog output full scale
N
OTA
7
:
U
,
V COMPONENTS
By selecting the U and V components, the speed value associated to the initial scale of the two
analog outputs is equal to the opposite of the speed value associated to the full scale of the
outputs.
For example, if the speed full scale value is set to 60 m/s, the speed range associated to the
analog outputs is -60…+60 m/s.
N
OTA
8
:
T
UNNEL MODE
The reference (arrow or notch) on the instrument housing should be aligned with the direction
of the tunnel.
The output 2 is set to full scale value if the wind blows in the direction pointed by the reference,
and to initial scale value if the wind blows in the opposite direction.
The initial scale of output 1 is associated to the speed value opposite to that associated to the
output full scale.
Instrument information:
Command
Reply
Description
G1
&VPnn.nn yyyy/mm/dd|
Version and date of firmware
RGS
&nnnnnnnn|
Serial number of instrument
RGI
&ccc…ccc|
User code
CGIccc…ccc
&|
Sets user code to ccc…ccc
(max. 34 characters)
Содержание DeltaOHM HD51.3D Series
Страница 39: ......