HD51.3D…
- 24 -
V2.0
Parameters for MODBUS-RTU mode:
Command
Reply
Description
CU5Annn
&|
Sets MODBUS address to nnn
The address should range within 1 and 247
Default
: 1
RU5A
& nnn|
Reads the MODBUS address setting
CU5Bn
&|
Sets the Baud Rate for MODBUS mode to:
9600 if n=3
19200 if n=4
38400 if n=5
57600 if n=6
115200 if n=7
Default
: 19200 (n=4)
RU5B
& n|
Reads Baud Rate setting for MODBUS
mode
CU5In
&|
Sets interface for MODBUS mode to:
RS232 if n=0
RS485 if n=1
RS422 if n=2
Default
: RS485 (n=1)
Note: with RS232 option you can connect to PC
or datalogger 1 instrument only; option useful
to do tests without RS232/RS485 conversion.
RU5I
& n|
Reads interface setting for MODBUS mode
CU5Mn
&|
Sets parity and stop bits for MODBUS
mode to:
8N1 if n=0 [No parity, 1 stop bit]
8N2 if n=1 [No parity, 2 stop bits]
8E1 if n=2 [Even parity, 1 stop bit]
8E2 if n=3 [Even parity, 2 stop bits]
8O1 if n=4 [Odd parity, 1 stop bit]
8O2 if n=5 [Odd parity, 2 stop bits]
The number of bits is fixed to 8
Default
: 8E1 (n=2)
RU5M
& n|
Reads the setting of parity and stop bits
for MODBUS mode
CU5Wn
&|
Sets waiting time after transmission in
MODBUS mode to:
Immediate reception if n=0
(violates protocol)
Waiting 3.5 characters if n=1
(respects protocol)
Default
: Waiting 3.5 characters (n=1)
RU5W
& n|
Reads current setting of waiting time after
transmission in MODBUS mode