5.2
E
LECTRICAL
C
ONNECTIONS
All connections are performed through a 19-pole M23 male connector situated at the bottom of
the instrument. The figure and the table below show numbers and function of the connector
contacts and the corresponding colors of the optional
CP51.x
cable wires:
Connector
pin number
CP51.x cable
wire color
Symbol
Description
1
White/Red
Not used
2
White/Grey
Not used
3
Yellow/Brown
RX +
Serial receive (input) positive
4
Brown/Green
HEAT -
Heater power supply negative
5
Violet
HEAT +
Heater power supply positive (24 Vdc)
6
Brown
HEAT -
Heater power supply negative
7
Grey/Brown
HEAT +
Heater power supply positive (24 Vdc)
8
Yellow
GND
Serial ground (isolated from V –)
9
Grey
TX -
Serial transmission (output) negative
“DATA –” main RS485 output
10
White/Yellow
AUX_B
“DATA +” auxiliary RS485 output (D+)
(*)
11
White
AUX_A
“DATA –” auxiliary RS485 output (D-)
(*)
12
Black
V -
Instrument power supply negative
13
Green
RX -
Serial receive (input) negative
14
Pink/Brown
OUT 1
Analog output 1 positive
15
Blue
GND
Analog ground (isolated from V –)
16
Red/Blue
OUT 2
Analog output 2 positive
17
White/Green
TX +
Serial transmission (output) positive
“DATA +” main RS485 output
18
--
Not used
19
Red
V +
Instrument power supply positive (12…30 Vdc)
--
Grey/Red
SHIELD
Cable shield
(*)