Confirm with ENTER. The following screen will appear:
PROBE CALIBRATION
CALIBRATION CO
1) Cal zero
2) Set sensitivity
▼▲
select
<ESC> exit/cancel
<ENTER> confirm
Confirm with ENTER. The following screen will appear:
PROBE CALIBRATION
CALIBRATION CO
SENSITIVITY SET
Sens 50 nA/ppm
CO 0 ppm
▼▲
set
<ESC> exit/cancel
<ENTER> confirm
Use the arrows
▼▲
to change the CO sensor sensitivity value and confirm using
ENTER.
If necessary, perform the CO sensor calibration at zero ppm.
RH
C
ALIBRATION
This procedure applies to
P37AB147
and to
Temperature
and
Humidity
combined
SICRAM probes.
Before starting the calibration operation, you should
check
if you need a new
calibration using the saturated solutions at 75.4 %RH and 33 %RH. You should
proceed with the calibration, only if you find an error in one of the two above
calibration points.
The calibration procedure removes the previous calibration data.
For a correct calibration of the sensor,
the first point should be 75 %RH
and the
second point 33 %RH.
Use the arrows
▼▲
to select the
3) Calibration RH
item:
PROBE CALIBRATION
1) Calibration CO
2
2) Calibration CO
3) Calibration RH
▼▲
select
<ESC> exit/cancel
<ENTER> confirm