![GHM DeltaOHM HD37AB1347 Скачать руководство пользователя страница 39](http://html.mh-extra.com/html/ghm/deltaohm-hd37ab1347/deltaohm-hd37ab1347_operating-manual_2220417039.webp)
HD37AB1347
- 39 -
V3.0
CO zero calibration with nitrogen bottle:
•
Using a screwdriver, open the
front small door of the probe.
•
Connect the tube from the
nitrogen with the rubber cap
on the CO sensor head.
Use the arrows
▲▼
to select the
2) Calibration CO
item:
PROBE CALIBRATION
1) Calibration CO
2
2) Calibration CO
3) Calibration RH
▼▲
select
<ESC> exit/cancel
<ENTER> confirm
Confirm with ENTER. The following screen will appear:
PROBE CALIBRATION
CO CALIBRATION
1) Cal zero
2) Sensitivity set
▼▲
select
<ESC> exit/cancel
<ENTER> confirm
Confirm with ENTER. The following screen will appear:
PROBE CALIBRATION
CO CALIBRATION
CAL ZERO
Cal zero 0 ppm
<ESC> exit/cancel
<ENTER> confirm
Содержание DeltaOHM HD37AB1347
Страница 1: ...Operating manual IAQ MONITOR HD37AB1347 www deltaohm com English Keep for future reference ...
Страница 5: ...HD37AB1347 5 V3 0 ...
Страница 63: ......