HD37AB1347
- 11 -
V3.0
O
PERATION
Before turning the instrument on, connect the SICRAM probes to the inputs:
8-pole male DIN 45326 connectors
, located in the upper part of the instrument.
You can connect the following probes to the inputs:
Input 1
Indoor Air Quality
for the
SICRAM
probes:
•
P37AB147: It measures CO
2
Carbon Dioxide, CO Carbon Monoxide, Relative
Humidity RH, Temperature T, and Atmospheric Pressure Patm.
•
Humidity
and
Temperature
combined probes.
•
Temperature
probes
with Pt100 sensor
.
Input 2
Temp-Air Velocity
for the
SICRAM
probes:
•
Hot-Wire Sensor Wind Speed
probes.
•
Vane Wind Speed
probes.
•
Temperature
probes
with Pt100 sensor
.
NOTE:
Connect the probes when the instrument is off. If a new probe is
connected and the instrument is on, it won’t be detected. You need to turn
the instrument off and on.
NOTE:
If you connect two temperature probes with Pt100 sensor to the two
inputs, only the Indoor Air Quality input probe will be detected. The Temp –
Air Velocity input probe will be ignored.
If a probe is disconnected when the instrument is on, you will get an acoustic signal
(one beep per second) and an indication on the display relevant to the physical
quantity being disconnected. The
LOST
message will be displayed.
During turning on, the following message is displayed for about 10 seconds:
Model HD37AB1347
Indoor Air Quality
Firm.Ver.=01.00
In addition to the
Delta OHM
logo,
the
instrument code and the firmware
version
are displayed.