HD37AB1347
- 36 -
V3.0
C
ALIBRATION
The instruments and probes are calibrated in the factory; no calibration is usually
required by the user. However, you can perform a new calibration.
With the HD37AB1347 instrument, you can perform the calibration of the following
connected probes:
•
With the
P37AB147
SICRAM probe connected to input 1 Indoor Air Quality,
you can perform the calibration of the RH (Relative Humidity), CO (Carbon
Monoxide), and CO
2
(Carbon Dioxide) sensors.
•
With the
temperature
and
humidity
combined SICRAM probes connected to
input 1 Indoor Air Quality, you can perform the calibration of the RH (Relative
Humidity) sensor.
No calibration is allowed for the Temperature sensor, and for the Hot-Wire
and Vane sensors.
To calibrate the probes correctly, a knowledge of and abiding by the physical
phenomena on which the measurement is based is fundamental: this is the reason
why it is recommended to abide by what is reported below carefully and only to
perform new calibrations if technically proficient.
PROBES CALIBRATION
1) Calibration CO
2
2) Calibration CO
3) Calibration RH
▼▲
select
<ESC> exit/cancel
<ENTER> confirm
CO
2
C
ALIBRATION
Ensure the
P37AB147
SICRAM probe is connected to input 1, Indoor Air Quality.
Use the arrows
▲▼
to select the
1) Calibration CO
2
item:
PROBES CALIBRATION
1) Calibration CO
2
2) Calibration CO
3) Calibration RH
▼▲
select
<ESC> exit/cancel
<ENTER> confirm