BAROsense
- 17 -
V1.2
Address
Description
Format
10
Maximum pressure value for the current analog output.
For the value, see
Note 3
below.
Default
: 110000 (=1100.00 hPa)
32-bit Integer
(2)
13
Minimum pressure value for the voltage analog output.
For the value, see
Note 3
below.
Default
: 60000 (=600.00 hPa)
32-bit Integer
(2)
15
Maximum pressure value for the voltage analog output.
For the value, see
Note 3
below.
Default
: 110000 (=1100.00 hPa)
32-bit Integer
(2)
Note 2
: The value is a 32-bit integer; two consecutive registers starting from the one indicated
must be accessed to read it. The register with lower address contains the least significant bits.
Note 3
: The value is an integer, expressed in the set unit of measurement and must take into
account the resolution corresponding to the set unit of measurement (
see the column “Modbus
protocol IR 0” of the table on page 5
). For example, if the set unit of measurement is hPa, the
value indicates hundredths of hPa (60000 equals 600.00 hPa. The number is automatically
converted if the unit of measurement is changed (which must therefore be set first).