HD52.3D…
- 29 -
V2.2
Command
Reply
Description
RAH
& n|
Reads the value corresponding to the wind
speed analog output full scale
N
OTA
5
:
U
,
V COMPONENTS
By selecting the U and V components, the speed value associated to the initial scale of the two
analog outputs is equal to the opposite of the speed value associated to the full scale of the
outputs.
For example, if the speed full scale value is set to 60 m/s, the speed range associated to the
analog outputs is -60…+60 m/s.
N
OTA
6
:
T
UNNEL MODE
In tunnel mode the wind direction measurement is not compensated by the magnetic compass,
but the measurement is referred to the direction indicated by the arrow on the instrument case.
The output 2 is set to full scale value if the wind blows in the direction of the arrow, and to initial
scale value if the wind blows in the opposite direction.
The initial scale of output 1 is associated to the speed value opposite to that associated to the
output full scale.
The arrow on the instrument case should be aligned with the direction of the tunnel.
Instrument information
Command
Reply
Description
G1
&Vnn.nn yyyy/mm/dd|
Version and date of firmware
RGD
&yyyy/mm/dd hh.mm.ss| Date and time of calibration
RGS
&nnnnnnnn|
Serial number of instrument
RGI &ccc…ccc|
User
code
CGIccc…ccc
&|
Sets user code to ccc…ccc
(max. 34 characters)