HD52.3D…
- 21 -
V2.2
Parameters for proprietary RS232 and RS485 modes:
Command
Reply
Description
CU1Ac
&|
Sets the address for proprietary RS485
mode to c value
The address is an alphanumeric character
ranging within 0…9, a…z, A…Z
Default : 0
RU1A
& c|
Reads the address for proprietary RS485
mode set in the instrument
CU1Bn
&|
Sets Baud Rate for proprietary RS485
mode to:
9600 if n=3
19200 if n=4
38400 if n=5
57600 if n=6
115200 if n=7
Default : 115200 (n=7)
RU1B
& n|
Reads Baud Rate setting for proprietary
RS485 mode
CU2Bn
&|
Sets Baud Rate for proprietary RS232
mode to:
9600 if n=3
19200 if n=4
38400 if n=5
57600 if n=6
Default : 57600 (n=6)
RU2B
& n|
Reads Baud Rate setting for proprietary
RS232 mode
CU1Dccccccccccc
&|
Sets measurements order in the string
sent in proprietary RS232 and RS485
modes
In the sequence ccccccccccc (max. 11
characters) each character identifies a
measurement according to the following
correspondence:
0
Barometric Pressure
1
Temperature (sensor Pt100)
2
Relative Humidity
3
Pyranometer
6
Wind speed (coordinates U,V)
7
Wind speed (intensity)
8
Wind Direction (Azimuth)
T
Sonic Temperature
C
Compass
E
Errors
note: the Wind Gust measurement and the
rainfall measurement are not available in
proprietary RS232 and RS485 modes
Default : 78
(see Note 2)
RU1D
& ccccccccccc|
Reads measurements order in the string
sent in proprietary RS232 and RS485 modes