HD52.3D…
- 14 -
V2.2
5.2
E
LECTRICAL
C
ONNECTIONS
All connections are performed through a 19-pole M23 male connector situated at the bottom of
the instrument. The figure and the table below show numbers and function of the connector
contacts and the corresponding wires of the optional CP52.x cable:
TAB. 5.A – Electrical connections
Connector
pin N
o
CP52.x cable
wire N
o
/color
Symbol
Description
1
Not
used
2
2 / Blue
RX SDI
Data line for SDI-12 connection
3
3 / Yellow
RX +
Serial receive (input) positive
4
HEAT -
Heater power supply negative
5
HEAT +
Heater power supply positive (10…30 Vdc)
6
6 / Pink
HEAT -
Heater power supply negative
7
7 / Violet
HEAT +
Heater power supply positive (10…30 Vdc)
8
8 / Grey
GND
Serial ground
9
9 / White
TX -
Serial transmission (output) negative
“DATA –” RS485 output
10
Not
used
11
Not
used
12
12 / Black
V -
Instrument power supply negative
13
4 / Grey-Pink
RX -
Serial receive (input) negative
14
10 / Brown
OUT 1
Analog output 1 positive
15
GND
Analog ground (isolated from V –)
16
11 / Green
OUT 2
Analog output 2 positive
17
5 / Red-Blue
TX +
Serial transmission (output) positive
“DATA +” RS485 output
18
V +
Instrument power supply positive (10…30 Vdc)
19
1 / Red
V +
Instrument power supply positive (10…30 Vdc)
--
Black (thick wire)
SHIELD
Cable shield
ATTENTION: The optional 12-pole CP52.x cable wire number does not always coin-
cide with the M23 connector pin number.
Connector
Reference
Visible side
male
connector