HD50CR-AS
- 16 -
V1.1
F900 submenu
(CLK MENU)
Clock settings.
F901
: year
F902
: month
F903
: day
F904
: hour
F905
: minutes
F906
: enabling of the automatic synchronization of the internal clock with a reference
server {On/OFF, default=OFF}
F907
: time zone {default=0}
FA00 submenu
(PSW MENU)
Password settings.
FA01
: exits the menu and deactivates immediately the password {YES/nO}. The
password will not remain active for some minutes as it normally happens
when exiting the menu: you will need to re-enter the password even if you
re-access immediately the menu.
FA02
: setting of the user-level password {default=0000}
FA03
: locking of the instrument keyboard {YES/nO, default=nO}.
If the
keyboard
is
locked, keep the
MENU/
↵
key pressed for 7 seconds to unlock it.
Fb00 submenu
(CAL MENU)
Calibration settings.
Fb03
: relative humidity sensor calibration at 75%RH
Fb04
: relative humidity sensor calibration at 33%RH
Fb05
: differential pressure calibration at zero
Fb06
: enabling of the differential pressure auto-calibration at zero {On/OFF, de-
fault=On}
Fb07
: interval, in minutes, of the differential pressure auto-calibration at zero {5…360,
default=60}
Fb08
: differential pressure second calibration point (in Pa)
Fb09
: reset of the differential pressure second calibration point
Fb10
: type of calibration used {FACT=factory / USEr=user, default=FACT}
Содержание Delta OHM HD50CR-AS
Страница 25: ...HD50CR AS 25 V1 1 Dimensions mm 8 218 151 53 30 70 234 92 8 76 1 5 44 3 1 2 3 3 x4 9 ESC MENU FUNC...
Страница 28: ...HD50CR AS 28 V1 1 NOTES...
Страница 29: ...HD50CR AS 29 V1 1 NOTES...
Страница 30: ...HD50CR AS 30 V1 1 NOTES...
Страница 31: ......