22
Some of the most frequent problems and their possible solutions are reported below.
Wrong pH measurement
. Carry out the following checks:
Check that the diaphragm is not obstructed and possibly clean it using the HD62PT solution.
Check that the reference system is not contaminated and, in case of a filling type electrode, replace
the electrolyte with the
KCL3M
solution (electrodes KP63, KP64 e KP90) or the
PROTELYTE
solution (electrodes KP61, KP71 e KP80).
Check that no air bubbles are present in the electrode tip and that it is sufficiently immersed.
Even dirt residuals deposited on the membrane can alter the measurement: use the
HD62PP
solu-
tion for protein cleaning.
Slow response or wrong measurements
. Possible causes are aging or erosion of the membrane or
a connector short circuiting.
Storage.
Keep the electrode immersed in the
HD62SC
solution.