![GHM Delta OHM HD3405.2 Скачать руководство пользователя страница 35](http://html1.mh-extra.com/html/ghm/delta-ohm-hd3405-2/delta-ohm-hd3405-2_manual_2220405035.webp)
- -
35
USB interface
1.1 - 2.0 electrically isolated
Connections
Temperature probe input
with SICRAM module or TP47 module
8-pole male DIN45326 connector
pH/mV input
Female BNC
Serial interface and USB
pole MiniDin connector
Mains adapter (cod. SWD10)
2-pole connector (positive at centre)
12Vdc/1A
Measurement of pH by Instrument
Measurement range
-2.000…+19.999pH
Resolution
0.01 or 0.001pH selectable from menu
Accuracy
±
0.001pH
±
1digit
Input impedance
>10
12
Ω
Calibration error @25°C
|Offset| > 20mV
Slope > 63mV/pH or Slope < 50mV/pH
Sensitivity > 106.5% or Sensitivity < 85%
Automatic/manual temperature compensation
-50…+150°C
Standard solutions automatically detected (@25°C)
1.679pH - 2.000pH - 4.000pH - 4.008pH -
4.010pH - 6.860pH - 6.865pH - 7.000pH -
7.413pH - 7.648pH - 9.180pH - 9.210pH -
10.010pH
Measurement of mV by Instrument
Measurement range
-1999.9…+1999.9mV
Resolution 0.1mV
Accuracy
±
0.1mV
±
1digit
Drift after 1 year
0.5mV/year
Measurement of temperature by Instrument
Pt100 measurement range
-200…+650°C
Pt1000 measurement range
-200…+650°C
Ni1000 measurement range
-50…+250°C
Resolution 0.1°C
Accuracy
±
0.1°C
±
1digit
Drift after 1 year
0.1°C/year
EMC standard regulations
Security
EN61000-4-2, EN61010-1 level 3
Electrostatic discharge
EN61000-4-2 level 3
Electric fast transients
EN61000-4-4 level 3,
EN61000-4-5 level 3
Voltage variations
EN61000-4-11
Electromagnetic interference susceptibility IEC1000-4-3
Electromagnetic interference emission
EN55020 class B