HD2013
- 17 -
V1.6
Resolution
Command
Description
PWD
BW
n
Sets the rain gauge resolution (amount of rainfall corresponding to an
emptying of the tipping bucket) of index
n
:
n
=0
0.1 mm
n
=1
0.2 mm
n
=2
0.25 mm
n
=3
0.5 mm
n
=4
1.0 mm
n
=5
0.254 mm (0.01 inches)
n
=6
0.508 mm (0.02 inches)
n
=7
1.016 mm (0.04 inches)
n
=X
custom value (set via the CCW command)
Reply of the instrument:
&| if the resolution is accepted
? if the index
n
is not correct
Yes
BR
Reads the rain gauge resolution.
The reply of the instrument is the index
n
of the resolution according to
the list given in the description of the editing command BW.
No
CCW
n…n
Sets
n…n
as custom value of the resolution. The value must be between
0.001 and 10.00 and is considered in the unit of measurement (mm or
inches) set in the instrument.
Example: CCW+0.42 sets 0.42 mm or 0.42 inches, depending on the
unit of measurement set, as custom value of the resolution.
Reply of the instrument:
&| if the value
n…n
is accepted
? if the value
n…n
is not correct
Warning
: the command CCW sets the value but not the use of the cus-
tom resolution; to use the cutom resolution, the command BWX shoul
be sent (see the command BW described above).
Yes
CCR
Reads the custom value of the resolution. The value is considered in the
unit of measurement (mm or inches) set in the instrument.
No
Output contact type
Command
Description
PWD
TNW
n
Sets the rain gauge output contact as normally open or normally closed:
n
=0
Normally OPEN,
n
=1
Normally CLOSED
Reply of the instrument:
&| if the setting is accepted
? if the index
n
is not correct
Yes
TNR
Reads the type of the rain gauge output contact.
Reply of the instrument:
0 if the contact is normally OPEN
1 if the contact is normally CLOSED
No