![GHM Delta OHM HD2013 Скачать руководство пользователя страница 26](http://html1.mh-extra.com/html/ghm/delta-ohm-hd2013/delta-ohm-hd2013_operating-manual_2220395026.webp)
HD2013
- 26 -
V1.6
Command
Instrument reply
Description
Amount of rainfall in the previous day
aC3!
atttn<CR><LF>
with: ttt = number of seconds necessary for the in-
strument to make the measure available (3
characters)
n = number of detected variables (1 character)
Note: ttt = 000 means that data are immediately
available.
Request of detection of the
amount of rainfall in the
previous day.
aD0!
a+n+r…r<CR><LF>
with:
n = number of emptying of the tipping bucket
r…r = amount of rainfall in the previous day in the
unit of measurement set in the instrument
(mm or inches)
Reads the amount of rain-
fall in the previous day.
Rainfall rate
aC4!
atttn<CR><LF>
with: ttt = number of seconds necessary for the in-
strument to make the measure available (3
characters)
n = number of detected variables (1 character)
Note: ttt = 000 means that data are immediately
available.
Request of detection of the
current rainfall rate.
aD0!
a+r…r<CR><LF>
with:
r…r = rainfall rate in the unit of measurement set in
the instrument (mm/h or in/h)
Reads the current rainfall
rate.
Power supply voltage and internal temperature
aC5!
atttn<CR><LF>
with: ttt = number of seconds necessary for the in-
strument to make the measures available
(3 characters)
n = number of detected variables (1 character)
Note: ttt = 000 means that data are immediately
available.
Request of detecting the
power supply voltage and
the internal temperature
of the rain gauge.
aD0!
a+v…v+t…t <CR><LF>
with: v…v = power supply voltage in V
t…t = internal temperature in the unit of
measurement set in the instrument (°C
or °F)
Reads the power supply
voltage and the internal
temperature of the rain
gauge.