Installation

 

 

1.

 

Choose a suitable location to mount the console, secure from unauthorized access to prevent tempering with the 

instrument, not too distant from the bore well and has easy reach to power socket.

 

2.

 

After finalizing the mounting location, check the availability of power by connecting a 12 volts 1A DC adapter

 

3.

 

Measure the length of cable to suspend into the bore well from the ground level and tie the cable firmly at the point 

where it leads into the bore well to avoid slipping.

 

4.

 

Next, insert the port at the end of the cable into the socket provided on the instrument body.

 

5.

 

Set sensor to “zero cal”.

 

6.

 

At this step, feed the length of cable measured in the instrument.

 

7.

 

Install the pressure transducer by lowering it into the well so that it is submerged below the water surface. Avoid 

dropping the transducer or permitting sharp contacts with the sides of the well casing. Do not allow the transducer 

to free fall into the well.

 

8.

 

The transducer should be installed at a point in the well that will not go dry. Estimate the lowest expected water 

level, and lower the transducer to the desired depth below the water level.

 