background image

Signet 

Conductivity/Resistivity

Transmitter

ENTER

62.50 uS/cm

25.0 C

Signet 8850-2 Conductivity/Resistivity Transmitter

2. Speci

fi

 cations

General

Compatible electrodes: Signet 3-28XX-1 Standard and Certi

fi

 ed 

(NIST) Series Conductivity/Resistivity Electrodes
Accuracy: 

±2% of reading

Enclosure:
• Case: 

PBT

• Panel gasket: 

Neoprene

•  Window: 

Polyurethane coated polycarbonate

•  Keypad: 

Sealed 4-key silicone rubber

•  Weight: 

Approx. 325 g (12 oz.)

Display:
•  Type: 

Alphanumeric 2 x 16 LCD

•  Contrast: 

User selected, 5 levels

•  Update rate: 

1.8 s

Electrical

Power: 12 to 24 VDC ±10%, regulated, 290 mA max.
Sensor input range: 
• Conductivity: 

0.01 

μ

S/cm to 400 000 

μ

S/cm

•  Resistivity:  

10 K

Ω

/cm to 100 M

Ω

/cm

•  TDS: 

0.023 to 200 000 PPM nominal

 (adjustable 

μ

S/PPM)

•  Temperature: 

PT 1000, -25 to 120 °C (-13 to 248 °F)

Measurements above 10 M

Ω

 (below 0.1 

μ

S) must be 

performed in solution temperatures from 20 to 100 °C.

4 to 20 mA Outputs:
•  Passive, isolated, fully adjustable and reversible 4 to 20 mA 

outputs are independently source selectable for conductivity or 
temperature.

•  Max loop impedance: 

  50 

Ω

 max. @ 12 V

  

325 

Ω

 max. @ 18 V

  

600 

Ω

 max. @ 24 V

•  Update rate: 

200 ms

•  Accuracy: 

±0.03 mA @ 25 °C, 24 V

Relay outputs (2 sets mechanical SPDT contacts):
•  Max. voltage rating: 

5 A @ 30 VDC or

 

5 A @ 250 VAC, resistive load

•  Hi or Lo programmable with adjustable hysteresis
•  Pulse programmable (maximum 400 pulses/minute)

Environmental

•  Operating temperature:  -10 to 70 °C (14 to 158 °F)
•  Storage temperature: 

-15 to 80 °C (5 to 176 °F)

•  Relative humidity: 

0 to 95%, non-condensing

•  Maximum altitude: 

2000 m (6562 ft)

•  Rating: 

NEMA 4X/IP65 front

• Insulation category: 

II

• Pollution degree: 

2

Standards and Approvals

•  CE, UL, CUL listed
• RoHS compliant
•  Manufactured under ISO 9001 and ISO 14001

CAUTION!

•  Remove power to unit before wiring input 

and output connections.

•  Follow instructions carefully to avoid 

personal injury.

Contents

1. Installation
2. Speci

fi

 cations

3. Electrical 

Connections

4. Menu 

Functions

1. Installation

ProcessPro transmitters are available in two styles: panel mount and 

fi

 eld mount. The panel mount is supplied with the necessary 

hardware to install the transmitter. This manual includes complete panel mounting instructions.
Field mounting requires a separate mounting kit. The 3-8050 Universal kit (159 000 184) enables the transmitter to be installed 
virtually anywhere. Detailed instructions for 

fi

 eld installation options are included with the 3-8050 Universal kit.

1.1 Panel Installation

  1.  The panel mount transmitter is designed for installation using a 1/4 DIN punch. For manual panel cutout, an adhesive template 

is provided as an installation guide. Recommended clearance on all sides between instruments is 1 inch.

  2.  Place gasket on instrument, and install in panel.
  3.  Slide mounting bracket over back of instrument until quick-clips snap into latches on side of instrument.
  4.  To remove, secure instrument temporarily with tape from front or grip from rear of instrument. DO NOT RELEASE.
 

  Press quick-clips outward and remove.

Output -

System Pwr
Loop -

System Pwr
Loop +

2

1

4

3

Sensr Gnd

(SHIELD)

Sensr IN

(RED)

Sensr V+

(BLACK)

7

6

5

   8050

106 mm (4.2 in.)

42 mm 

(1.7 in.)

64 mm 

(2.5 in.)

92 mm
(3.6 in.)

97 mm

(3.8 in.)

56 mm 
(2.2 in.)

41 mm 
(1.6 in.

)

Optional
Rear
Cover

96 mm

(3.8 in.)

96 mm
(3.8 in.)

quick-clips

gasket

panel

terminals

mounting

bracket

latch

Panel Mount
Installation Detail

82 mm
(3.23 in.)

SIDE VIEW
Field Mount

Panel Mount

SIDE VIEW

FRONT VIEW

Field Mount &

Panel Mount

3-8850.090-2  Rev. J  4/12 

English

*3-8850.090-2*

English

Chinese RoHS (Go to www.gfsignet.com for details)

Содержание Signet ProcessPro 8850-2

Страница 1: ...nder ISO 9001 and ISO 14001 CAUTION Remove power to unit before wiring input and output connections Follow instructions carefully to avoid personal injury Contents 1 Installation 2 Specifications 3 El...

Страница 2: ...val Procedure 1 Press the orange terminal lever downward with a small screwdriver to open terminal jaws 2 When fully open remove wire from terminal Wiring Tips Do not route sensor cable in conduit con...

Страница 3: ...Loop AUX Power AUX Power Transmitter 2 Terminals PLC Terminals Loop Input 4 20 mA Transmitter Terminals System Pwr Loop System Pwr Loop AUX Power AUX Power Power Supply 12 24 VDC Power Supply 12 24 VD...

Страница 4: ...keys The selections will scroll in a continuous loop Changing the VIEW display does not interrupt system operations No key code is necessary to change display selection Output settings cannot be edite...

Страница 5: ...2 The Key Code is keys in sequence After entering the Key Code the display will show the first item in the selected menu Step 3 Scroll menu with or arrow keys Step 4 Press key to select menu item to b...

Страница 6: ...ut Loop Range S 0 0000 100 00 Select the minimum and maximum values for the 4 to 20 mA current loop output Be sure to modify this setting if you change the Cond Units Relay1 Source Cond Select Conduct...

Страница 7: ...able display and output response Temp Display C Select temperature units C or F Temperature Comp 2 00 Set the percent change in Conductivity caused by a 1 C change in temperature May be from 0 00 to 1...

Страница 8: ...t value If the instrument does not display the correct value Temperature 0 5 C Conductivity 2 of reading service is required A Accuracy Verification with Precision Resistors Electronic Calibration 1 S...

Страница 9: ...ayed temperature and conductivity values in the spaces provided Displayed temperature T1 _______ C Displayed conductivity C1 _______ S 3 Cool the sample solution close to the minimum process temperatu...

Страница 10: ...condition if your process operates at or near the limits of the electrode range Check sensor for correct range Check Decimal setting in OPTIONS menu Check Calibrate menu settings for incompatible SOUR...