18
2581 FlowtraMag
®
Meter
App Con
fi
guration - Sensor Setup Continued
Information
Sensor information, Bluetooth Device Tag, Permanent
Totalizer, Resettable Totalizer, Calibration Adjustment
Factor, Zero O
ff
set Adjustment
Application Setup
To set Averaging, Sensitivity,
Low Flow Cut O
ff
, Position
of Flow, Flow Units and
Totalizer Unit.
Loop
Set 4 mA, 20 mA, Error
condition of the current output
alarm (3.6 or 22 mA), adjust
your 4 to 20 mA setting and
select output mode.
Loop adjustment is a live
update.
Then press
DONE
or
press on
GF logo
to get
back to re-connect screen.
Calibration
Custom Calibration of
Rate, Volumetric, Zero
Flow Calibration.
Note: You must press WRITE to save
your changes to the sensor.
Otherwise it saves to your phone only.
To switch between digital (S
3
L) and Frequency and/or 4 to 20
Active or Passive, use the GF Con
fi
g Tool App. On the loop
screen, use the drop down to select digital (S
3
L) or Freq
and/or Active or Passive 4 to 20 mA. Press WRITE after
making selection.
If the GF Con
fi
g Tool password has been lost or forgotten,
connect blue wire to white wire of the sensor output cable while
unit is powered (for 2 to 5 seconds.) Disconnect blue wire from
white wire after 5 seconds. Password will reset to factory original
(last 6 digits of serial number.)
To delete saved 2581 FlowtraMag in iOS:
Swipe right and select the trash can icon.
To delete saved 2581 FlowtraMag in Android:
Swipe right, "Are you sure you want to delete this
device?", choose Yes or No.
Technical Note
The totalizer in the 2581 FlowtraMag is independent from
the totalizer in the 9900/9950 transmitters. If the totalizer
is reset on the 2581 FlowtraMag, it does not reset the
totalizer on the 9900 or 9950.
iOS version
Android version