30
Site Requirements and Specifications
Floor
Maximum Load (re water)
300 kg / 661 lbs
Maximum floor load
5400 N/m
2
Maximum floor pressure
0,6 N/mm
2
Slope
1:50
Ceiling height
Min. 2510 mm + 50 mm
(98 3/4” + 2”)
Transport and Storage
Ambient temperature range
-20°C - 70°C
(-4°F - 158°F)
Relative humidity range
10% - 80% Rh
(not condensed)
Atmospheric pressure range
500 hPa to 1100 hPa
Water (Hot and Cold)
Connection
3/4” B.S.P female
Upper limit static pressure
6 bar / 87 PSI
Recommended dynamic pressure, at 8°C (46°F) CW
and at 80°C (176°F) HW flow of 58 L/min (15,3 Gal/min)
3 bar / 43 PSI
Lower limit dynamic pressure at 8°C CW and at 80°C HW
and 26 L/min flow (GB)
0,8 bar / 12 PSI
Lower limit dynamic pressure at 8°C (46°F) CW
and 80°C (176°F) HW degrees and 29 L/min
(7,7 Gal/min) flow.
(North America without disinfection)
1 bar / 15 PSI
Lower limit dynamic pressure at 8°C (46°F) CW
and 80°C (176°F) HW degrees and 45 L/min
(11,9 Gal/min) flow.
(All other markets) (North America with disinfection)
2 bar / 29 PSI
Maximum pressure difference between hot
and cold water
+/- 0,5 bar / +/- 7 PSI
Cold water lower limit temperature
2°C (36°F)
Cold water upper limit temperature
20°C (68°F)
Hot water lower limit temperature
60°C (140°F)
Hot water upper limit temperature
80°C (176°F)
Содержание Arjohuntleigh Parker
Страница 1: ...Parker Assembly and Installation Instructions with people in mind 06 AL 00_12GB 02 2016...
Страница 4: ...4 Intentionally left blank...
Страница 13: ...13 Intentionally left blank...
Страница 23: ...23 HOT COLD Red Blue 1 2...
Страница 32: ...32 Intentionally left blank...
Страница 33: ...33 Intentionally left blank...
Страница 34: ...34 Intentionally left blank...