22
Water Connection
•
•
•
Water supply
The water supply can be arranged according to
two main alternatives:
•
Water connection from floor.
•
Water connection from wall.
•
1. Connect water hoses (not included in
delivery) to the water inlets for hot and cold
water inlets in the wall or in the floor.
1. It is possible to turn the inlets sideways if
a better direction for the hoses is needed.
(Figure 1.)
2. Connect the other ends of the water hoses
to the hot /cold water inlets on the tub.
(Figure 2.)
•
•
In house requirement
Min. flow / supply pipe
25 l/min (6.6 US gal/min)
50 l/min CW + HW (13.2 US gal/min CW + HW)
Water connection hoses requirement
Flexible hoses
Nominal size
DN15
Connection dimension
3/4” B.S.P female
Connect with shut off valves
between unit and supply
Min. through Ø16 mm
Approved to local regulation for temperature / pressure
WARNING
Hot and cold water supply must be fit-
ted with local isolating valves.
CAUTION
All plumbing connections and installa-
tions must comply with all relevant
local, state or federal regulations and
plumbing codes.
NOTE
Flush connection hoses prior to connect-
ing to the mixer.