Repair and adjustment
12
Service instructions
5015108-00
Edition 0403
Calibration
Conductivity meter
Check the output signal from conductivity meter
To check the output signal, proceed as follows.
1. Press
[
on the conductivity meter.
2. Enter code 22 with
Y
and
Z
.
3. Press
[
. Setup 1 appears.
4. Choose with
Y
and
Z
Output.
5. Press
[
. Sel. Type appears.
6. Choose LIN with
Y
and
Z
.
7. Press
[
.
8. Choose 4-20 mA with
Y
and
Z
.
9. Press
[
.
10.Set with
Y
and
Z
so that 4 mA =0.000 µS/cm.
11.Press
[
.
12.Set with
Y
and
Z
so that 20 mA = 200 µS/cm.
13.Press
[
.
14.Press
Y
and
Z
at the same time to log out.
Set the cell constant
To set the cell constant, proceed as follows:
1. Press
[
.
2. Enter code 22 with
Y
and
Z
.
3. Press
[
. Setup 1 appears.
4. Press
[
until the display shows Cellconst.
5. With
Y
and
Z
, set the relevant cell constant; see the calibration certificate.
6. Press
[
.
7. Press
Y
and
Z
at the same time to log out.
Содержание 8666
Страница 1: ...SERVICE INSTRUCTIONS Washer disinfector Getinge 8666 8668 Mfg no SEV0320078 5015108 00 ...
Страница 2: ......
Страница 4: ......
Страница 6: ...Safetyregulations 2 Serviceinstructions 5015108 00 Edition 0403 ...
Страница 8: ......
Страница 10: ...Technicaldata 2 Serviceinstructions 5015108 00 Edition 0403 ...
Страница 12: ...Technicaldata 4 Serviceinstructions 5015108 00 Edition 0403 Alternativeconnectionarrangements ...
Страница 14: ......
Страница 24: ......
Страница 26: ......
Страница 33: ......
Страница 37: ......
Страница 39: ......
Страница 47: ......
Страница 49: ...Repairandadjustment 2 Serviceinstructions 5015108 00 Edition 0403 ...
Страница 95: ......
Страница 96: ......
Страница 97: ......
Страница 98: ......
Страница 99: ......
Страница 100: ......
Страница 101: ......
Страница 102: ......
Страница 103: ......
Страница 104: ......
Страница 105: ......
Страница 106: ......
Страница 107: ......
Страница 108: ......
Страница 109: ......
Страница 110: ......
Страница 111: ......
Страница 112: ......
Страница 113: ......
Страница 114: ......
Страница 115: ......
Страница 116: ......
Страница 117: ......
Страница 118: ......
Страница 119: ......
Страница 120: ......
Страница 121: ......
Страница 122: ......
Страница 123: ......
Страница 124: ......
Страница 125: ......
Страница 126: ......
Страница 127: ......
Страница 128: ......
Страница 129: ......
Страница 130: ......
Страница 131: ......
Страница 132: ......
Страница 133: ......
Страница 134: ......
Страница 135: ...This product is manufactured by GETINGE DISINFECTION AB Ljungadalsgatan 11 Box 1505 351 15 Växjö Sweden ...