8 Cleaning and disinfection
CAUTION
Do not use solvents, such as ether, acetone or paraffin
oil; such substances can damage the material of the
housing.
CAUTION
Before cleaning or disinfecting the device, remove the
battery and close the battery compartment.
Before performing surface disinfection, clean the device.
Use a lint-free cloth slightly moistened with a mild soap solution to
wipe the device.
Disinfect the device at regular intervals, prior to first use and be-
fore passing it on to another person.
GETEMED recommends to use a 70% alcohol solution for disin-
fection.
Please note the application time of the disinfectant (10 minutes
for 70% alcohol solution).
Subsequently use a lint-free cloth slightly moistened to remove
remainders of the disinfectant.
The device withstands up to 700 cleaning / disinfection cycles (7
years of normal use).
Re-processing by way of a machine is excluded for the product.
The product is not intended for sterilization.
Store re-processed products in a dust-free and dry place.
The instructions listed above have been validated by the medical
device manufacturer as being suitable for preparing the medical
device for reuse. It is the responsibility of the processor to ensure
that the processing actually carried out with the equipment, mate-
rials and personnel used in the processing facility achieves the
desired result. This requires verification and / or validation and
routine monitoring of the process.