GERATHERM Bodystik Скачать руководство пользователя страница 4

User Manual 

Bodystik 

BODYSTIK_MAN_ENG_V1.2.2_REV01.DOCX 

Page 4 

 

2.

 

Bodystik 

2.1

 

Description and Scope of Delivery 

 

The medical product Bodystik is a PC based Bodyplethysmograph, 
which is used in PFT measurement to determine the lung volumes as 
well as the resistance of the respiratory system. It could be used as a 
single device or in combination with other products of Geratherm 
Respiratory. 
The Bodystik consists of a pressure stable, fully glazed Aluminium 
cabin with integrated electronics. It is connected to a computer via 
the integrated USB interface. The established Ergoflow Flow sensor 
Principle is used for flow measurement.  
The Bodystik can only be used together with the tool and gear wagon 
authorized by Geratherm Respiratory. The galvanic isolation of the 
electric power system is done by the tool and gear wagon. All 
peripheral equipment like computers, printers or monitors should 
only be connected to the electric power system by the tool and gear 
wagon.  
The Software Blue Cherry is delivered with the Bodystik. Its Standard 
version contains Measurements for SVC, FVC, Bodyplethysmography 
as well as Pre/Post analysis and trend analysis. Furthermore it 
contains a powerful database for patient and test information and 
has a modular design which offers the possibility of combination with 
other measurements. 
The modular and flexible software concept allows user specific 
configuration of Blue Cherry. A User may add to the predefined 
medication list, select between different predicted equations sets 
and add further predicted equations, define new parameters, 
configure results table and change patient data input view.  
Blue Cherry can be connected to office information software via a 
GDT interface. There are different export possibilities into Excel and 
XML. Using the standard Microsoft ®clipboard, all charts may be 
imported into standard software such as MS Word or MS 
PowerPoint. 

Содержание Bodystik

Страница 1: ...Bodystik User Manual Version 1 2 2 Date 17th October 2014...

Страница 2: ...2 6 Cabin controls 11 2 2 7 Power on the Bodystik 12 2 3 Security 12 2 3 1 General Security information 12 2 3 1 1 Norms and Guidelines 13 2 3 1 2 Electromagnetic compatibility 13 2 3 2 Secure use of...

Страница 3: ...e integrated into even the largest of requirements This document contains important information for the operation of Blue Cherry and Bodystik We strongly recommend reading this manual carefully in ord...

Страница 4: ...ent like computers printers or monitors should only be connected to the electric power system by the tool and gear wagon The Software Blue Cherry is delivered with the Bodystik Its Standard version co...

Страница 5: ...Description Bodystik All over glazed aluminium cabin with electrical stand and height adjustable patient chair as well as tool and gear wagon with integrated isolation transformer Blue Cherry Softwar...

Страница 6: ...tion has to be established On request Geratherm Respiratory provides diagrams parts lists descriptions setting instructions as well as other documents which could help the operator to setup maintain a...

Страница 7: ...he bottom of the stand Compressed air connection for gas supply 1 sample gas connection 2 electric unit connection 3 2 2 2 Shutter Unit Now move the shutter 1 as far as it will go into the closing hea...

Страница 8: ...be connected additionally Therefore move the demand valve 1 as far as it will go into the closing head just like shown in picture 2 After that connect the junction cable of the demand valve with the j...

Страница 9: ...nt up to 0 3 CH4 up to 0 3 CO up to 0 3 C2H2 optionally 21 O2 remain N2 Open the gas bottle carefully after all connections has been established like described 2 2 4 Flow sensor installation Open the...

Страница 10: ...k of the Bodystik Now connect the Bodystik via the provided USB cable to a free USB port at your computer Connection panel of the Bodystik power supply jack 1 USB port 2 CO Diffusion analyzer 3 CO Dif...

Страница 11: ...ch were shown below are arranged out of the cabin Emergency stop switch 1 door status LED 2 glows green if door is closed door opener 3 push button stand up 4 push button stand down 5 microphone 6 The...

Страница 12: ...l is intended for medical technical personnel familiar with the performance of Bodyplethysmographic Testing procedures Users have to be familiar with the testing method and its clinical acceptation Th...

Страница 13: ...Guidelines The Geratherm Respiratory products are manufactured in accordance with DIN EN ISO 13485 According to the European directive 93 42 EEC Bodystik fulfils the requirements of annex II In additi...

Страница 14: ...ed signal connections The Bodystik should not be used in conjunction with multiple connection plug sockets or extension leads Bodystik must only be used with the tool and gear wagon provided by Gerath...

Страница 15: ...The flow sensor should be replaced in these conditions In addition extreme temperature as well as extreme altitude may impair the flow measurement Consider the site conditions indicated in the techni...

Страница 16: ...hutter To protect both user and subject from cross contamination we recommend using the Neumofilt bacteria and virus filter Thus avoids even the contamination of the measurement instrumentation Anyhow...

Страница 17: ...d the maximum allowed valued specified in the Norm DIN EN 60601 1 For some non medical devices a higher case leakage current is allowed Those higher limits can only be accepted outside patient area It...

Страница 18: ...in patient area and resource B in medical used room Resource A within patient area and resource B in medical used room See remarks below table Resource A within patient area and resource B in a non me...

Страница 19: ...nce with local directives 2 3 6 Product Labelling Example of typical product labelling SN designates the serial number for the device The first 2 digits define the manufacturing year The example 13 sh...

Страница 20: ...ation component type BF according to DIN EN 60601 1 Indicates single use device protection type II according to MDD Separate collection of electro and electronic devices IPX0 Indicates no protection a...

Страница 21: ...minor maintenance is required to ensure trouble free operation Interval Maintenance work Once a day Flow sensor flow sensor housing and shutter unit cleaning and disinfect Once a day Check O Rings fo...

Страница 22: ...be provided by Geratherm Respiratory or an authorized service partner of Geratherm Respiratory Geratherm Respiratory will not accept warranty claims on any unauthorised repairs 3 2 Warranty exclusion...

Страница 23: ...25 Hz 3 or 20ml s Volume Measurement range Accuracy 0 20L 3 or 50mL Cabin pressure Measurement principle Measurement range Accuracy Resolution Semiconductor 0 25 kPa 1 5 FS 24 bit Mouth pressure Measu...

Отзывы: