User Manual
Bodystik
BODYSTIK_MAN_ENG_V1.2.2_REV01.DOCX
Page 14
2.3.2
Secure use of Bodystik
The Bodystik is a medical device to perform PFT measurements. It is
highly recommend that the user observes all warning notices in order
to ensure a safe use of the device. Following parts of the device are
usable in the patient environment:
-
Cabin
-
Medical cart
-
Electric unit
-
Shutter unit
-
Demand valve and gas analyser (only with CO-Diffusion
option)
Equipment which is not part of the measurement
system described above, should not be connected.
Only the mentioned devices should be connected to the
described signal connections.
The Bodystik should not be used in conjunction with
multiple connection plug sockets or extension leads.
Bodystik must only be used with the tool and gear
wagon provided by Geratherm Respiratory. The tool and
gear wagons multiple connection plug socket may be
loaded with up to 1000 VA. Only equipment, which is
part of the measurement system, should be connected.
Bodystik must only be used in combination with
peripheral devices which fulfill the Norm DIN EN 60950-
1 for office devices or the norm DIN EN 60601-1 for
medical electrical devices. Always ensure the minimum
requirements for the computer are met (see technical
specification). Non-Medical devices should not be used
close to a patient always ensure a minimum distance of
1.5m between patient and non-medical computer