Instructions for Use
Spirostik Complete
Serial no.: xx|8|102|yyy and 2103xxxxx
Page 78
Version: 7 | Release date: 14. July 2021
12.4
Electromagnetic Compatibility / EMC Guidelines
The manufacturer tests his products for emitted interference and
interference resistance. Compliance with the relevant standards
and directives is certified in the Declaration of Conformity that
accompanies this device. The results of the EMC test can be
found in the following chapter.
12.4.1
Emitted Interference Guideline and Manufacturer Declaration
Guidelines and manufacturer’s declaration – electromagnetic emissions
The Spirostik Complete is determined for operation in an electromagnetic environment as
specified below. The user / operator of the Spirostik Complete should ensure that it is operated
in this environment.
Measurement of
electromagnetic emissions
Compliance
Electromagnetic environment – Guideline
RF emissions CISPR 11
Group 1
The Spirostik Complete uses RF energy
exclusively for its internal function. Therefore, its
RF emission is very low and is improbable to
interfere with adjacent electronic equipment.
RF emissions CISPR 11
Class B
The Spirostik Complete is suitable for use in all
establishments, including living areas and those
directly connected to the public supply network,
which also supplies buildings used for residential
purposes.
Emissions of harmonic
oscillations according to
IEC 61000-3-2
Class A
Emissions of voltage
fluctuations / flicker
according to
IEC 61000-3-3
Compliance