Instructions for Use
Spirostik Complete
Serial no.: xx|8|102|yyy and 2103xxxxx
Page 62
Version: 7 | Release date: 14. July 2021
8.2.1
Checking for Damage
8.2.1.1
General
All parts of Spirostik Complete should be checked for visible
mechanical damage (cracks, tears) each day. If damage is
ascertained, the corresponding component must be replaced.
For replacing components see:
•
Chap. 4.1.4 “System Construction and Electrical
Safety”
•
Chap. 6.2 “Returning to Operation after Servicing /
Cleaning Work”
8.2.1.2
Checking / Replacing the Handle
During the daily inspection of the sealing rubber of the Spirostik
Complete handle, attention must be paid to mechanical damage
or signs of abrasion. The double tube must be firmly connected
to the sealing rubber on the handle with no kinks or other
damage. If necessary, the handle must be replaced completely.
The Hhndle must generally be replaced once a year. Proceed as
described in chap. 6.1.2.1 "Connect the Handle".