Instructions for Use
Spirostik Blue
Serial no.: 2111xxxxx
Version: 7 | Release date: 02. August 2021
Page 73
residual moisture from being
sucked in.
•
Repeat the previous steps at the
second hole for pressure
measurement.
6.
Visual inspection
No contamination and residues of
cleaning agents and disinfectant must
be visible.
9.2.3
Drying Bellows
The drying bellows can be cleaned of dirt using a cleaning
solution / weak soapy water.
The following disinfection options are permitted:
•
Disinfection bath (aqueous solution with, for example,
ammonium chloride; do not use alcohol-based
disinfectants, oils, fats, solvents or acids).
•
Brief boiling in plenty of boiling water (< 3 min) - longer
temperatures above 80 °C damage the material.
•
Gas sterilisation possible.
Information on the application and dosing of the disinfectant
can be found in the separate manual of the respective
manufacturer. The safety instructions contained must be
observed.
To remove and insert the flow sensors, please read:
•
Chap. 7.7 “Removing the Flow Sensor”
•
Chap. 7.4 “Inserting the Flow Sensor”