Independent of this, the following should also be observed:
Possible danger to life.
Reason: Electric shock. Misdiagnosis. Therefore:
•
Do not modify or use the Spirostik Blue or
complete system contrary to the respective
manufacturer's specification!
Reason: Electric shock. Therefore:
•
Under no circumstances use or connect any
devices, systems, equipment and other products
that are not part of the complete system! This
also includes only components approved by the
manufacturer such as Bluetooth dongle,
(rechargeable) batteries, battery chargers,
(connection) cables and computers that are not
permitted by the manufacturer. See chap. 12
"Technical Specifications".
•
Ensure that in the patient environment
−
a
distance of 1.5 meters to the patient
−
there are
no accessible electrical parts (interfaces, plugs,
etc.) that are not isolated from the mains with an
isolation voltage of 4 kV!
•
When connecting Spirostik Blue to other
devices, make sure that all connections and
cables are secured and cannot come loose by
themselves!
•
Do not connect any portable multiple sockets or
extension cables to the system!