Attention
Spirostik Blue could be damaged.
Reason: Improperly treated components. Therefore:
•
For cleaning and disinfection only use those active
substances that are approved by the manufacturer (see
chap. 2.2.2 “Consumable Items / Auxiliary Materials”)!
•
Follow the instructions on the concentration and dwell
time stated by the cleanser and disinfectant
manufacturer!
•
Do not place Spirostik Blue in the cleaning and
disinfection solutions!
•
The device contains electrical components that will be
damaged by doing so!
•
Remove the (rechargeable) batteries from Spirostik Blue
before cleaning and disinfection work!
•
Wipe off any remaining moisture with a dry cloth!
•
Observe the maximum permissible temperature of
75 °C for cleaning and disinfection!
•
Do not dry with heat!
Possible impairment of the measuring function.
Reason: Moisture in the connection piece. Therefore:
•
Ensure that there is no moisture in the
connection piece before recommissioning!