T R O U B L E S H O O T I N G
Error Messages on Display
Main Display Problem
Possible Solution
OvEr
•
Conductivity is
>200.0 mS
•
Salinity is > 70.00 ppt
•
Completely submerge
the probe.
•
Allow sufficient time for
the electrode and Temp
probe stabilization.
•
Recalibrate with
correct value for the
conductivity standard.
•
Replace conductivity
standard.
•
Clean cell.
•
Return for service.
OvEr/Undr
during
calibration
•
Cell Constant Calibration
is out of range
Main
Display
Secondary
Display
Problem
Possible Solution
OvEr/Undr
OvEr
Temperature >90.0 °C
• Decrease/
Increase the sample
temperature.
• Return for service.
Undr
Temperature
< -10.0 °C
Opening the Data File with Excel
®
Depending on the region and language setting of your PC,
measurement data might be formatted incorrectly by Excel
®
when the
data file is opened.
This is sometimes encountered with German set as the PC language,
as a German date format typically utilizes a decimal (DD.MM.YYYY).
The EC300M utilizes a decimal as the radix, so a temperature of
31.1
is sometimes interpreted by Excel
®
as
31. Jan
when German is set as
the PC language.
If a data file is opened in Excel
®
and measurement data is incorrectly
interpreted as something other than a number, please follow these
steps:
1. Open a blank Excel
®
spreadsheet.
2. Go to the
Data
tab and select
From Text
.