8

Note

: Be certain to press 

Enter 

 

 to reset the cell constant 

to 5.00. If 

MODE

 is pressed, the unit retains the previous cell 

constant and calibrates from a value that is already offset.

Temperature Coefficient

The unit uses the temperature coefficient to calculate temperature 
compensated conductivity. The default value is 1.91%. To change the 
temperature coefficient, use the up and down keys to adjust the value 
between 0 and 4.00%. Press 

Enter

 

 to save the new value, or press 

MODE

 to cancel the change and display the Temperature Reference 

screen.

Temperature Reference

The unit uses the temperature reference value to calculate temperature 
compensated conductivity. The default value is 25°C. To change the 
temperature coefficient, use the up and down keys to adjust the value 
between 15 and 25°C. Press 

Enter

 

 to save the new value, or press 

MODE

 to cancel the change and display the Conductivity Calibration 

screen.

Conductivity Calibration

1.  Immerse the probe in a standard of known conductivity, preferably 

a standard in the middle range of the solutions to be measured. 
Completely submerge the probe without touching the sides of the 
calibration container. Shake the probe lightly to remove any air 
bubbles trapped in the conductivity cell.

2.  Allow temperature to stabilize. The message ‘rAng’ (range) may 

display briefly to indicate unit auto-ranging; this is normal. After 
temperature stabilization, use the up and down keys to adjust the 
conductivity value to that of the conductivity standard at 25°C. Press 

Enter

 

 to calibrate. The unit beeps twice to indicate a successful 

calibration, then automatically switches to normal operation mode.

Conductivity Measurements

1.  Turn the unit on. Place the probe in the solution to be measured. 

Completely submerge the probe. Shake the probe lightly to 
remove any trapped air bubbles in the conductivity cell.

2.  Press 

MODE

 to enter the desired measurement mode. The 

message ‘rAng’ (range) may appear briefly on the display indicate 
auto-ranging; this is normal. Allow temperature to stabilize before 
taking measurements.

606042REF-B-EC300A-EC300M-Manual-English.indd   8

12/11/2017   10:07:43 AM

Содержание EcoSense EC300A

Страница 1: ...USER MANUAL English Fran ais Espa ol Portugu s EcoSense EC300A EcoSense EC300M Portable Conductivity Salinity and Temperature Instruments User Manual DOCUMENT 606042REF...

Страница 2: ...amage or failure caused by i failure to install operate or use the product in accordance with YSI s written instructions ii abuse or misuse of the product iii failure to maintain the product in accord...

Страница 3: ...t Modes 6 Calibration 7 TDS 7 CELL 7 Temperature Coefficient 8 Temperature Reference 8 Conductivity Calibration 8 Conductivity Measurements 8 Saving Viewing and Deleting Data 9 Downloading Data to a C...

Страница 4: ...Accessories Part Numbers 14 Technical Services 14 Specifications 14 606042REF B EC300A EC300M Manual English indd 4 12 11 2017 10 07 39 AM...

Страница 5: ...l contaminants that may affect probe connections 3 Wait for unit and all connections to dry before reinstalling the battery and resuming operation 4 If the unit does not function correctly after step...

Страница 6: ...attery cover and o ring Figure 1 Replace the 9V battery Replace the battery cover and o ring be sure to align the o ring correctly to prevent a bad seal and fasten the two battery cover screws Battery...

Страница 7: ...vailable with a single four electrode cell Other features include automatic conductivity ranging automatic temperature compensation long battery life and 50 60 Hz AC noise rejection This meter is univ...

Страница 8: ...ction indicator EC300M only 8 Displays when measuring salinity 9 Calibration mode indicator 10 Data set number 11 Temperature display 12 Indicates TDS measurement 13 Indicates conductivity measurement...

Страница 9: ...oSense Mode CAL Recal Timer Date Time Up and down arrow keys Increases or decreases the display value as desired 3 DO200M INPUT EcoSense Mode CAL Recal Timer Date Time Calibration key Press to enter t...

Страница 10: ...confirms calibration steps and confirms data deletion On the EC300M this key confirms recal timer entry and date time selections 8 YSI EcoSense Handheld Overlay 12 19 2016 DO200M colors Black backgrou...

Страница 11: ...quentially display the following sections Note Press Enter to accept any values changes in each section and automatically advance to the next section If there are no changes the unit accepts the curre...

Страница 12: ...in a standard of known conductivity preferably a standard in the middle range of the solutions to be measured Completely submerge the probe without touching the sides of the calibration container Shak...

Страница 13: ...ice to return to the measurement mode If Each is selected the Up and Down arrow keys will allow you to scroll through the saved data records Press Enter to delete the selected record All records after...

Страница 14: ...urn to the run screen Any value between 0 and 60 days can be selected Set the value to 0 to disable the Recal Timer Date Time Settings EC300M Only A short press i e key is not held of the Date Time ke...

Страница 15: ...for service Undr Temperature 10 0 C Opening the Data File with Excel Depending on the region and language setting of your PC measurement data might be formatted incorrectly by Excel when the data fil...

Страница 16: ...t Wizard choose Delimited 5 Under Step 2 of the Text Import Wizard choose Semicolon 6 Under Step 3 click on the column with the incorrectly formatted data This column should be highlighted in black Ch...

Страница 17: ...1 Dip the cell in cleaning solution and agitate for two to three minutes Any foaming acid tile cleaner such as Dow Chemical Bathroom Cleaner should clean adequately For a stronger cleaner use a soluti...

Страница 18: ...com S P E C I F I C AT I O N S These specifications represent typical performance and are subject to change without notice For the latest product specification information please visit YSI s website a...

Страница 19: ...ure 0 to 50 C 32 to 122 F Operating Range Relative Humidity Up to 95 Instrument Casing Waterproof IP 67 with cable connector cap installed Weight with battery 272 grams 6 lbs Dimensions L x W x H 18 7...

Страница 20: ...2013 IEC 61326 1 2012 IEC 61000 3 2 2005 IEC 61000 3 3 2008 IEC 61000 4 2 2008 IEC 61000 4 3 2006 IEC 61000 4 4 2004 IEC 61000 4 6 2008 IEC 61000 4 8 2009 Item 606042REF Revision B December 2017 For t...

Отзывы: