key (EC300M 

only). Press to enter the Recal 
Timer input display. 

5

Mode

 key. Selects display mode. 

In Normal operation, press Mode 
to switch the display between 
uncompensated conductivity, 
temperature compensated 
conductivity, salinity, total dissolved 
solids (TDS), Recall and Delete. In 
calibration mode, this key exits the 
current calibration and displays 
the next calibration parameter. 

7

YSI EcoSense Handheld Overlay

12/19/2016

 

DO200M

colors:

Black background

PMS 7482 C power symbol

PMS 130 C logo

Reverse white all other imagery

DO

200

M

INPUT

EcoSense

®

Mode

CAL

Recal

Timer

Date

Time

Enter 

key. Pressing Enter saves 

the current measurement 
into memory, confirms mode 
selection (rcl/delete), confirms 
calibration steps, and confirms 
data deletion. On the EC300M, 
this key confirms recal timer 
entry and date/time selections.

8

YSI EcoSense Handheld Overlay

12/19/2016

 

DO200M

colors:

Black background

PMS 7482 C power symbol

PMS 130 C logo

Reverse white all other imagery

DO

200

M

INPUT

EcoSense

®

Mode

CAL

Recal

Timer

Date

Time

Date/Time 

key (EC300M only). A 

short press (i.e. key is not held) of 
the Date/Time key changes the 
display in the upper right corner 
to be either Date or Time. Pressing 
and holding for 3 seconds will allow 
date and time information to be 
updated. Pressing and holding for 6 
seconds will allow for the date (e.g. 
MM/DD/YYYY) and time format (12 
or 24 hour) settings to be changed. 

O P E R A T I O N A L   P R O C E D U R E S

Measurement Modes

1.  Temperature - Current solution temperature continually displays.

2.  Temperature Compensated Conductivity - Measurement of 

conductivity, compensated to 25°C or another specified value 
between 15 and 25°C. Expressed as uS/cm or mS/cm with a 
flashing “°C”.

606042REF-B-EC300A-EC300M-Manual-English.indd   6

12/11/2017   10:07:43 AM

Содержание EcoSense EC300A

Страница 1: ...USER MANUAL English Fran ais Espa ol Portugu s EcoSense EC300A EcoSense EC300M Portable Conductivity Salinity and Temperature Instruments User Manual DOCUMENT 606042REF...

Страница 2: ...amage or failure caused by i failure to install operate or use the product in accordance with YSI s written instructions ii abuse or misuse of the product iii failure to maintain the product in accord...

Страница 3: ...t Modes 6 Calibration 7 TDS 7 CELL 7 Temperature Coefficient 8 Temperature Reference 8 Conductivity Calibration 8 Conductivity Measurements 8 Saving Viewing and Deleting Data 9 Downloading Data to a C...

Страница 4: ...Accessories Part Numbers 14 Technical Services 14 Specifications 14 606042REF B EC300A EC300M Manual English indd 4 12 11 2017 10 07 39 AM...

Страница 5: ...l contaminants that may affect probe connections 3 Wait for unit and all connections to dry before reinstalling the battery and resuming operation 4 If the unit does not function correctly after step...

Страница 6: ...attery cover and o ring Figure 1 Replace the 9V battery Replace the battery cover and o ring be sure to align the o ring correctly to prevent a bad seal and fasten the two battery cover screws Battery...

Страница 7: ...vailable with a single four electrode cell Other features include automatic conductivity ranging automatic temperature compensation long battery life and 50 60 Hz AC noise rejection This meter is univ...

Страница 8: ...ction indicator EC300M only 8 Displays when measuring salinity 9 Calibration mode indicator 10 Data set number 11 Temperature display 12 Indicates TDS measurement 13 Indicates conductivity measurement...

Страница 9: ...oSense Mode CAL Recal Timer Date Time Up and down arrow keys Increases or decreases the display value as desired 3 DO200M INPUT EcoSense Mode CAL Recal Timer Date Time Calibration key Press to enter t...

Страница 10: ...confirms calibration steps and confirms data deletion On the EC300M this key confirms recal timer entry and date time selections 8 YSI EcoSense Handheld Overlay 12 19 2016 DO200M colors Black backgrou...

Страница 11: ...quentially display the following sections Note Press Enter to accept any values changes in each section and automatically advance to the next section If there are no changes the unit accepts the curre...

Страница 12: ...in a standard of known conductivity preferably a standard in the middle range of the solutions to be measured Completely submerge the probe without touching the sides of the calibration container Shak...

Страница 13: ...ice to return to the measurement mode If Each is selected the Up and Down arrow keys will allow you to scroll through the saved data records Press Enter to delete the selected record All records after...

Страница 14: ...urn to the run screen Any value between 0 and 60 days can be selected Set the value to 0 to disable the Recal Timer Date Time Settings EC300M Only A short press i e key is not held of the Date Time ke...

Страница 15: ...for service Undr Temperature 10 0 C Opening the Data File with Excel Depending on the region and language setting of your PC measurement data might be formatted incorrectly by Excel when the data fil...

Страница 16: ...t Wizard choose Delimited 5 Under Step 2 of the Text Import Wizard choose Semicolon 6 Under Step 3 click on the column with the incorrectly formatted data This column should be highlighted in black Ch...

Страница 17: ...1 Dip the cell in cleaning solution and agitate for two to three minutes Any foaming acid tile cleaner such as Dow Chemical Bathroom Cleaner should clean adequately For a stronger cleaner use a soluti...

Страница 18: ...com S P E C I F I C AT I O N S These specifications represent typical performance and are subject to change without notice For the latest product specification information please visit YSI s website a...

Страница 19: ...ure 0 to 50 C 32 to 122 F Operating Range Relative Humidity Up to 95 Instrument Casing Waterproof IP 67 with cable connector cap installed Weight with battery 272 grams 6 lbs Dimensions L x W x H 18 7...

Страница 20: ...2013 IEC 61326 1 2012 IEC 61000 3 2 2005 IEC 61000 3 3 2008 IEC 61000 4 2 2008 IEC 61000 4 3 2006 IEC 61000 4 4 2004 IEC 61000 4 6 2008 IEC 61000 4 8 2009 Item 606042REF Revision B December 2017 For t...