19
APPENDIX C. SWAGELOK TUBE FITTING INSTRUCTIONS
These instructions apply to
one inch (25 mm) and smaller fittings
.
C.1 Installation
1)
Fully insert the tube into the fitting until it bumps against the shoulder.
Figure 11 - Tube Insertion
2)
Rotate the nut until it is finger-tight. (For high-pressure applications as well as high-safety-
factor systems, further tighten the nut until the tube will not turn by hand or move axially in
the fitting.)
3)
Mark the nut at the six o’clock position.
Figure 12 - Make a Mark at Six O’clock
4)
While holding the fitting body steady, tighten the nut one and one quarter turns, until the
mark is at the nine o’clock position. (Note: For 1/16", 1/8", 3/16", and 2, 3, and 4 mm
fittings, tighten the nut three-quarters of a turn until the mark is at the three o’clock position.)
Figure 13 - Tighten One and One-Quarter Turns
Содержание 4500HT
Страница 2: ......
Страница 4: ......
Страница 15: ...9 Figure 5 A Typical Model 4700 VW Temperature Sensor Calibration Report ...
Страница 16: ...10 Figure 6 Typical Calibration Report at 25 C ...
Страница 17: ...11 Figure 7 Typical Calibration Report at 100 C ...
Страница 18: ...12 Figure 8 Typical Calibration Report at 150 C ...
Страница 19: ...13 Figure 9 Typical Calibration Report at 200 C ...
Страница 20: ...14 Figure 10 Typical Calibration Report at 250 C ...