16
APPENDIX A. SPECIFICATIONS
A.1 4500HT Piezometer
Standard Ranges ¹
kPa
2
: −100 to 700
MPa
2
: 1, 2, 3, 5, 7.5, 10, 20, 50, 75, 100, 150
Resolution
0.025% F.S. (minimum)
Linearity
< 0.5% F.S. (±0.1% F.S. optional)
Accuracy
3
±0.1% F.S.
Overrange
1.5 × Rated Pressure
Temperature Range
0 °C to +250 °C
Thermal Zero Shift
<0.05% F.S./ °C
Diaphragm
Displacement
<0.001 cm
3
at F.S.
Dimensions (L × ø)
4
191 × 19 mm (7.5" × 0.75")
For pressure ranges of 100 to 700 kPa, and 1, 2, 3, 5, 7.5, 10 MPa
Frequency Range
1400-3500 Hz
Table 2 - HT Vibrating Wire Piezometer Specifications
Accuracy of Geokon test apparatus: 0.1%
Contact Geokon for specific application information.
Notes:
¹ Other ranges available on request.
2
PSI = kPa × 0.14503 or MPa × 145.03
3
Accuracy established under laboratory conditions.
4
Please contact Geokon for dimensions of ranges higher than 10 MPa.
A.2 Mineral Insulated Cable
Conductors
4-conductors, 22 AWG, solid copper
Sheath
Stainless Steel
Sheath Wall
0.76 mm (0.03")
Nominal O.D.
4.76 mm (0.1875")
Coil ID
1 m / 3 ft.
Table 3 - Mineral Insulated Cable Specifications
A.3 Tubular Encapsulated Cable (TEC)
Conductors
4-conductors, 24 AWG, stranded, tinned, copper
Insulation
PFA
Sheath
316L Stainless Steel
Nominal O.D.
4 mm / 0.160"
Sheath Wall
0.76 mm / 0.030"
Coil Diamter
1 m / 3 ft.
Table 4 - Tubular Encapsulated Cable Specifications
Содержание 4500HT
Страница 2: ......
Страница 4: ......
Страница 15: ...9 Figure 5 A Typical Model 4700 VW Temperature Sensor Calibration Report ...
Страница 16: ...10 Figure 6 Typical Calibration Report at 25 C ...
Страница 17: ...11 Figure 7 Typical Calibration Report at 100 C ...
Страница 18: ...12 Figure 8 Typical Calibration Report at 150 C ...
Страница 19: ...13 Figure 9 Typical Calibration Report at 200 C ...
Страница 20: ...14 Figure 10 Typical Calibration Report at 250 C ...